iii

 IM DLM5058-03EN

Checking the Contents of the Package

Unpack the box, and check the following before operating the instrument. If the wrong items have 
been delivered, if items are missing, or if there is a problem with the appearance of the items, 
contact your nearest YOKOGAWA dealer.

DLM5034, DLM5038, DLM5054, or DLM5058

Check that the product that you received is what you ordered by referring to the model name and 
suffix code given on the name plate on the left side panel.

 

Model

Suffix Code

1

Specifications

DLM5034

DLM5034 Mixed Signal Oscilloscope  

4 ch 16-bit logic, 350 MHz, 12.5 M/50 M/125 Mpoints

DLM5038

DLM5038 Mixed Signal Oscilloscope  

8 ch 16-bit logic, 350 MHz, 12.5 M/50 M/125 Mpoints

DLM5054

DLM5054 Mixed Signal Oscilloscope  

4 ch 16-bit logic, 500 MHz, 12.5 M/50 M/125 Mpoints

DLM5058

DLM5058 Mixed Signal Oscilloscope  

8 ch 16-bit logic, 500 MHz, 12.5 M/50 M/125 Mpoints

Power cord

2

-D

UL/CSA standard, rated voltage: 125 V

-F

VDE standard, Korean standard, rated voltage: 250 V

-R

Australian standard, rated voltage: 250 V

-Q

BS standard, rated voltage: 250 V

-H

Chinese standard, rated voltage: 250 V

-N

Brazilian standard, rated voltage: 250 V

-T

Taiwanese standard, rated voltage: 125 V

-B

Indian standard, rated voltage: 250 V

-U

IEC plug Type B, rated voltage: 250 V

-Y

No power cord

3

Language

(One option can 

be selected for 

the factory default 

message language 

and panel sheet 

language.)

-HJ

Japanese

-HE

English

-HC

Chinese

-HG

German

-HF

French

-HK

Korean

-HL

Italian

-HS

Spanish

1   For products whose suffix code contains “Z,” an exclusive manual may be included. Please read it along 

with the standard manual.

2   Make sure that the attached power cord meets the designated standards of the country and area that you 

are using it in.

3   Prepare a power cord that complies with the standard specified by the country or region that the instrument 

will be used in.

Содержание DLM5034

Страница 1: ...User s Manual DLM5034 DLM5038 DLM5054 DLM5058 Mixed Signal Oscilloscope Getting Started Guide IM DLM5058 03EN 4th Edition...

Страница 2: ...w us to serve you best by completing the user registration form accessible from our website https tmi yokogawa com support Contact Us If you want to resolve a technical support issue or need to contac...

Страница 3: ...ve any questions or find any errors please contact your nearest YOKOGAWA dealer Copying or reproducing all or any part of the contents of this manual without the permission of YOKOGAWA is strictly pro...

Страница 4: ...m Preparation to Displaying Waveforms Measuring Waveforms and Saving Screen Captures DLM5034 DLM5038 DLM5054 DLM5058 Mixed Signal Oscilloscope Communication Interface User s Manual IM DLM5058 17EN The...

Страница 5: ...ignal Oscilloscope 8 ch analog 16 bit logic 500 MHz 12 5 M 50 M 125 Mpoints Power cord2 D UL CSA standard rated voltage 125 V F VDE standard Korean standard rated voltage 250 V R Australian standard r...

Страница 6: ...01949 miniature passive probes instead of the four standard 701937 passive probes 8ch model only 1 For products whose suffix code contains Z an exclusive manual may be included Please read it along wi...

Страница 7: ...na IM 00C01C01 01Z1 1 Safety manual European languages PIM 113 01Z2 1 List of worldwide contacts Manual CD A1033US 1 For the types of manuals that CD contains see the next page Standard accessories ar...

Страница 8: ...be affixed to the front panel protection cover Hook and loop fasteners are attached to the back side of the soft case Soft case Front panel protection cover Hook and loop fasteners Where to affix the...

Страница 9: ...range 702907 1 DC to 200 MHz bandwidth maximum input voltage 1000 V DC AC peak operating temperature range 40 C to 85 C excluding the phase adjustment side IM 702907 01EN PBL100 logic probe1 701988 1...

Страница 10: ...voltage 1000 Vrms FET probe 700939 1 DC to 900 MHz bandwidth maximum input voltage 10 V DC AC peak IM 700939 01E Differential probe 701977 1 DC to 50 MHz bandwidth maximum differential input voltage 7...

Страница 11: ...12 0 5 V IM 701934 01E Deskew signal source 701936 1 Output voltage square wave approx 0 to 5 V output current square wave approx 0 to 100 mA small current approx 0 to 1 A large current IM 701936 01EN...

Страница 12: ...uctions In the manual the symbol is used in conjunction with the word WARNING or CAUTION WARNING Calls attention to actions or conditions that could cause serious or fatal injury to the user and preca...

Страница 13: ...bols are used on this instrument Handle with care Refer to the user s manual or service manual This symbol appears on dangerous locations on the instrument which require special instructions for prope...

Страница 14: ...ding Further do not use this power cord with other instruments Connect the Protective Ground Terminal Be sure to connect the protective earth to prevent electric shock before turning ON the power The...

Страница 15: ...als is Other O Do not use it to measure the main power supply or for Measurement Categories II III and IV Install or Use the Instrument in Appropriate Locations Do not install or use the instrument ou...

Страница 16: ...de terre Avant de mettre sous tension veiller brancher la mise la terre de protection afin d viter les chocs lectriques Le cordon d alimentation que vous utilisez pour l instrument est un cordon trois...

Страница 17: ...surer l alimentation lectrique ni pour les cat gories de mesure II III et IV Installer et utiliser l instrument aux emplacements appropri s Ne pas installer ni utiliser l instrument l ext rieur ou dan...

Страница 18: ...Regulation valid only in the EEA and UK Battery Regulation in the UK Batteries are included in this product This marking indicates they shall be sorted out and collected as ordained in the EU battery...

Страница 19: ...Connecting Logic Probes 2 19 2 7 Affixing the Panel Sheet 2 22 2 8 Loading Roll Paper Into the Built In Printer Optional 2 23 Chapter 3 Common Operations 3 1 Key and Jog Shuttle Operations 3 1 3 2 To...

Страница 20: ...Chapter 6 Specifications 6 1 Signal Input Section 6 1 6 2 Triggering Section 6 4 6 3 Time Axis 6 8 6 4 Display 6 8 6 5 Features 6 9 6 6 Built in Printer Option 6 17 6 7 Storage 6 17 6 8 USB Ports for...

Страница 21: ...B mouse USB storage device or USB printer USB keyboard or mouse sec 3 4 USB storage device or printer see the user s manual Keys and knobs sec 1 2 LOGIC C and D logic signal input ports Connect logic...

Страница 22: ...2 3 Main power switch sec 2 3 Inlet holes Left Side Panel GP IB port optional Use this connector to communicate with the instrument through the GP IB interface Communication Interface User s Manual I...

Страница 23: ...ector Use this connector to view the instrument s display on an external display sec 4 3 External trigger input terminal USB port for PCs Use to connect to a PC that has a USB port Communication Inter...

Страница 24: ...any of the CH1 to CH8 keys to specify the waveform that you want to adjust the position of This knob has a push switch You can press the knob to reset the trigger position to 0 00 div You can also pr...

Страница 25: ...push switch You can press the knob to reset the trigger position to 50 DELAY Key When you press the DELAY key the key illuminates Then you can set the trigger delay using the POSITION knob You can re...

Страница 26: ...y Press SHIFT and then press MODE to display an action on trigger menu or a GO NO GO menu B TRIG Key Use this key to set trigger combinations with the Edge or Enhanced trigger and to set the trigger B...

Страница 27: ...a menu for setting the signal acquisition mode RUN STOP Key Starts and stops signal acquisition according to the trigger mode The key illuminates while the instrument is acquiring signals SINGLE Key P...

Страница 28: ...want to control This knob has a push switch If you press the knob so that Fine illuminates the resolution is set to fine mode SEARCH Key Displays a waveform search menu SHIFT SEARCH SERIAL BUS key Pr...

Страница 29: ...e related tasks UTILITY Key Press this key to display a menu for calibrating the instrument connecting to a network or PC running a self test and setting the date and time menu language message langua...

Страница 30: ...at you selected using the jog shuttle If there are two items on the jog shuttle setting menu you can press the key to change the item that the jog shuttle controls Move the SET key left and right to m...

Страница 31: ...and horizontal SCALE LEVEL and ZOOM Push the knobs straight If you push a knob at an angle it may not operate properly If this happens push the knob straight one more time CAUTION Do not push a knob s...

Страница 32: ...ver range is occurring Appears when the input coupling is set to AC Appears when the input coupling is set to DC Appears when the input coupling is set to DC50 Appears when the bandwidth limit is not...

Страница 33: ...r point the pre trigger section is being acquired Post Trig Data after the trigger point the post trigger section is being acquired Acquisition Mode Normal Normal mode Norm Hi Res Normal mode with hig...

Страница 34: ...rection zoom range of Z2 Normal waveform Time div Z2 display record length Zoom vertically When the vertical zoom factor of the selected waveform is not x1 00 the vertical zoom amount is displayed in...

Страница 35: ...stogram Waveform display area Area for displaying the result of statistical processing Measured waveform parameter values statistics If the setup menu is not displayed measured waveform parameter valu...

Страница 36: ...e using the instrument in a location where the power supply specifications are different purchase a power cord that matches the specifications of the region that the instrument will be used in Operati...

Страница 37: ...nd the case When you carry the instrument be careful not to get your hand caught between the wall installation surface or other objects and the instrument French AVERTISSEMENT Lorsque vous attrapez ou...

Страница 38: ...instrument l ext rieur ou dans des endroits expos s la pluie ou l eau Installer l instrument de mani re pourvoir imm diatement le d brancher du secteur en cas de fonctionnement anormal ou dangereux A...

Страница 39: ...ce above and beyond the space shown in the figure above to carry out the procedure Ambient Temperature and Humidity Use the instrument in the following environment Ambient temperature 5 C to 40 C Ambi...

Страница 40: ...e legs or stand push it out until it locks into place To put away the stand push it inward while returning it to its original position Stand Movable legs Tilted installation Horizontal installation WA...

Страница 41: ...suivantes Ne manipuler les pieds ou le support escamotable que lorsque l instrument est plac sur une surface stable Ne pas manipuler les pieds ou le support escamotable lorsque l instrument est inclin...

Страница 42: ...the instrument will not be grounded If there is no AC outlet that is compatible with the power cord that you will be using and you cannot ground the instrument do not use the instrument French AVERTIS...

Страница 43: ...z Permitted supply frequency range 48 Hz to 63 Hz Maximum power consumption 290 VA This instrument can use a 100 V or a 200 V power supply The maximum rated voltage differs according to the type of po...

Страница 44: ...he Power Is Turned On Turn off the power switch and check the following items Check that the power cord is securely connected Check that the correct voltage is coming to the power outlet See section 2...

Страница 45: ...tch on the right side panel while the power switch on the front panel is turned on the setup data immediately before the power switch is turned off will not be stored properly The next time you turn o...

Страница 46: ...ut voltage level Over range indication CAUTION The probe interface terminals are located near the input terminals on this instrument Do not short the probe interface terminals When you connect a probe...

Страница 47: ...de enti re de la forme d onde s affiche sur l afficheur et v rifier le niveau de tension d entr e D passement de plage ATTENTION Sur cet instrument les bornes d interface de sonde se situent proximit...

Страница 48: ...and approximately 16 pF parallel or 50 1 0 Notes about Connections When connecting a probe to the instrument for the first time perform phase correction of the probe as described in section 2 5 Correc...

Страница 49: ...0 1 2 Bandwidth DC to 500 MHz within 3 dB Rise time 700 ps or less typical Maximum input voltage 400 Vrms Typical value represents a typical or average value It is not strictly warranted Precautions t...

Страница 50: ...on page vii CAUTION Do not use the probe power terminals option on the instrument s right side panel for purposes other than supplying power to an FET probe current probe or differential probe Be sure...

Страница 51: ...probe example of characteristics Measured current and current consumption of the 701929 current probe example of characteristics Current consumption mA 50 40 30 20 10 0 10 20 30 40 50 Positive power...

Страница 52: ...ger le circuit interne Procedure 1 Turn on the power switch 2 Connect the probe to the input terminal to which the signal is to be applied 3 Connect the tip of the probe to the signal output terminal...

Страница 53: ...signals on the probe To correct the phase you must adjust this trimmer so that flat frequency characteristics are obtained Be sure to correct the phase of a probe that you are using for the first time...

Страница 54: ...e For information about derating based on frequency see the user s manual of the logic probe that you are using The eight input lines on the logic port have a common ground In addition the ground for...

Страница 55: ...diff rentes Cela pourrait endommager l instrument les sondes logiques connect es et les appareils connect s Pour toute information sur la manipulation de la sonde logique se reporter au manuel d utili...

Страница 56: ...GS s pulse interpolation for logic waveforms Threshold level 40 V resolution 0 05 V 6 V resolution 0 05 V Threshold accuracy1 0 1 V 3 of setting Minimum input voltage 500 mVp p 300 mVp p Input impeda...

Страница 57: ...can affix the panel sheet over the panel sheet that is affixed to the instrument when it is shipped from the factory If the operation panel is stained wipe gently with a soft dry clean cloth Make sure...

Страница 58: ...as quickly as possible after you break its protective seal If you attach a plastic film that contains plasticizing material such as vinyl chloride film or cellophane tape to the paper for a long time...

Страница 59: ...ection du coupe papier l extr mit du cache de l imprimante Vous pourriez vous blesser Raise the handle Printer cover OPEN COVER Button Roll paper cutter section Print head Align the ends of the roll p...

Страница 60: ...Use the jog shuttle and the SET key to configure the settings D Pressing the soft key selects the item that you can control using the jog shuttle The jog shuttle setting menu shows the selected soft k...

Страница 61: ...If you press RESET when you are using the jog shuttle to set a value or select an item the setting is reset to its default value depending on the operating state of the instrument the setting may not...

Страница 62: ...the appropriate item 3 Press SET The operation varies depending on the selected item Switches the selected item Selects or Displays a selection menu Selects the item that the cursor is on A check box...

Страница 63: ...dialog boxes Drag Drag refers to the act of pressing your finger against the screen and moving your finger Dragging is used to select a waveform cursor zoom box or the like and move it Pinch Out and...

Страница 64: ...con A menu of the instrument s front panel keys the top menu appears 2 Tap the item you want to select The item that you tapped is executed or a setup menu for the tapped item appears at the right sid...

Страница 65: ...enu you can set the value as follows When there is a single icon tapping inside the frame displays a numeric keypad or selection menu When a numeric keypad appears tap the numeric keypad to input a va...

Страница 66: ...o select The operation varies depending on the selected item Switches the selected item Selects or Displays a selection menu Select an item A check box is selected or cleared Displays a setup dialog b...

Страница 67: ...to select The operation varies depending on the selected item Select an item Select a file folder or drive Scroll bar Drag to scroll through the file list Closes the file list To cancel the selection...

Страница 68: ...on the window you controlling as follows Control Main window Zoom window Tap1 Selects the waveform Selects the waveform Drag horizontally2 Changes the timescale zoom position4 Pinch horizontally Chan...

Страница 69: ...m position and zoom factor of all waveforms are reset to default values Note This is the same function as Vertical Zoom Reset on the ZOOM menu Turning off the Zoom Window Display 1 Hold down the zoom...

Страница 70: ...ion Vertical direction Displayed zoom window Zoom1 Horizontal ZOOM1 Zoom1 Horiz Vert ZOOM1 Zoom2 Horizontal ZOOM2 Zoom2 Horiz Vert ZOOM2 The zoom area set in step 2 is displayed in the zoom window The...

Страница 71: ...tle Select the appropriate item using the soft keys and change the value using the jog shuttle and the SET key This manual sometimes describes this operation simply as using the jog shuttle Note For i...

Страница 72: ...the cursor 2 Press SET to enter the character If a character string has already been entered use the arrow soft keys to move the cursor to the position you want to insert characters into Use the CAPS...

Страница 73: ...A connector of the USB keyboard to the instrument You can connect or remove the USB cable regardless of whether the instrument s power switch is on or off hot plugging is supported When the power swi...

Страница 74: ...USB Mouse Devices You can use mouse devices that are compliant with USB HID Class Version 1 1 Note For USB mouse devices that have been tested for compatibility contact your nearest YOKOGAWA dealer C...

Страница 75: ...s the top menu appears 2 Click the item that you want to select The item that you clicked is executed or a setup menu for the clicked item appears at the right side of the screen The top menu disappea...

Страница 76: ...the value as follows When there is a single icon clicking inside the frame displays a numeric keypad or selection menu When a numeric keypad appears click the numeric keypad to input a value Click En...

Страница 77: ...Displays a selection menu Select an item A check box is selected or cleared Displays a setup dialog box Switches the selected item Selects or Closes the setup dialog box Pointer Scroll Operation If a...

Страница 78: ...iv and Time div Controlling the Vertical Scale V div When the waveform of a channel whose voltage is being measured appears on the display move the pointer close to the V div value Rotate the mouse wh...

Страница 79: ...ntrolling Time div during Zooming During zoom display move the pointer close to the Time div value displayed in the zoom window Rotate the mouse wheel forward to increase the zoom factor and rotate it...

Страница 80: ...or measured data and files The instrument is factory shipped with a given date and time You must set the clock before you start measurements Procedure Press UTILITY and then the System Configuration s...

Страница 81: ...e set Hour to 9 and Minute to 0 Checking the Standard Time Using one of the methods below check the standard time of the region where you are using the instrument Check the Date Time Language and Regi...

Страница 82: ...ill be 0 V Source Channels CH1 to CH8 CH1 to CH4 on 4ch models Logic waveforms are displayed with the same settings that were used before you executed auto setup Waveforms Displayed before the Executi...

Страница 83: ...to the maximum record length at which the instrument can acquire waveforms repeatedly Mode Normal Hi Resolution OFF Sampling Mode Interpolation Trigger Settings Trigger Type Edge Trigger Mode Auto Hol...

Страница 84: ...Cannot Be Reset to Their Factory Default Values Date and time settings Communication settings Language setting English or Japanese Measured value font size setting Transparent display on and off tran...

Страница 85: ...changes Execute calibration in a stable temperature environment ranging from 5 to 40 C 23 5 C recommended Do not apply signals when calibrating Calibration may not be executed properly when input sig...

Страница 86: ...een When the RUN STOP key or SINGLE key is not illuminated waveform acquisition is stopped Stopped appears in the upper right of the screen Instrument Operation When the Acquisition Mode Is Set to Ave...

Страница 87: ...SET key left and right Moving Cursors and Scrolling To scroll through the screen or move the cursor in the table of contents or index turn the jog shuttle Moving to the Link Destination To move to a...

Страница 88: ...Input Terminal TRIGGER IN This terminal is used when an external signal is used as the trigger source Item Specifications Connector type BNC Maximum input voltage 40 V DC ACpeak or 28 Vrms At 1 MHz an...

Страница 89: ...signal is output when the instrument triggers The signal level is normally high but goes low when a trigger occurs Item Specifications Connector type BNC Output level 3 3 V CMOS Output impedance Appro...

Страница 90: ...me 2 1 1 High high level 3 period The sum of the pre trigger time and the internal processing time The minimum period is 50 ns 2 Low low level 3 period The sum of the post trigger time and the interna...

Страница 91: ...t circuit et ne pas y appliquer une tension externe Si vous le faites l instrument risque de ne pas fonctionner correctement Video Signal Output Connector VIDEO OUT You can use video signal output to...

Страница 92: ...e monitor using an analog RGB cable 3 Turn on the instrument and the monitor Note An RGB video signal is always running through the VIDEO OUT connector The monitor display may flicker if you place the...

Страница 93: ...on result is NO GO the output signal level the TTL level temporarily changes from high level H to low level L GO OUT Signal When the determination result is GO the output signal level the TTL level te...

Страница 94: ...l de d termination GO NO GO un autre appareil veiller ne pas connecter la mauvaise broche de signal Cela pourrait endommager cet instrument ou l instrument connect Ne pas brancher de c ble USB sur le...

Страница 95: ...et ne pas lui appliquer de tension externe Si vous le faites l instrument pourrait mal fonctionner Utiliser un c ble de connexion 701982 01 701982 02 Ne utiliser aucun autre c ble Si vous le faites l...

Страница 96: ...f the target instrument Connect the cable securely to the synchronous operation I O connector and check that it is locked in place Note For details on the synchronous operation DLMsync feature see the...

Страница 97: ...e not correct Confirm that the ambient temperature and humidity are within their specified ranges 2 2 After turning on the power wait at least 30 minutes to allow the instrument to warm up 2 3 Calibra...

Страница 98: ...gain after changing settings or reacquiring waveforms 13 The corresponding field was not found 14 Action on trigger is completed 2 27 15 The instrument is set to remote mode by the communication contr...

Страница 99: ...can check the synchronization status with the following icons displayed at the top of the screen Finding the master blinking Time synchronization being established Time synchronization established 1...

Страница 100: ...ed current record size Change the compressed record size and execute again 533 Assigned path does not exist Check the network setting and configuration Chapter 18 534 Assigned path does not exist Chec...

Страница 101: ...Chapter 22 Features Guide IM DLM5058 01EN Printer Errors Code Message and Corrective Action Section 550 Printer error Confirm the printer status 551 Cannot detect printer Turn ON the printer Check con...

Страница 102: ...the cycle Chapter 9 665 Search target data does not exist Execute search after analysis is completed 666 Improper action setting The saved data type is either Waveform group or Analysis group This ca...

Страница 103: ...e for the execution conditions of auto deskew Chapter 222 1 User s Manual IM DLM5058 02EN 2 Features Guide IM DLM5058 01EN Setting Errors Code Message and Corrective Action Section 800 Improper Date T...

Страница 104: ...service is required 907 Calibration failure Set V div to the highest sensitivity and turn the coarse adjustment trimmer of the current probe so that the signal is within 2 division from the center of...

Страница 105: ...This section explains how to test whether the instrument s memory keyboard and printer are working properly Procedure 1 Press UTILITY and then the Self Test soft key The following menu appears Test ty...

Страница 106: ...ard Test Set the test type to Keyboard Executes the test Soft keyboard test Printer Test Set the test type to Printer Executes the test Synchronous Operation Feature Test Set the test type to DLMsync...

Страница 107: ...if the print density is correct If an error occurs the print density will not be correct DLMsync Tests whether or not the option s synchronous operation feature is operating properly Connect the two...

Страница 108: ...following screen appears Explanation Displayed Information Model The model number Record Length The record length Sample Rate The maximum sample rate Band Width The frequency bandwidth Serial No The s...

Страница 109: ...ur nearest YOKOGAWA dealer When making a purchase please indicate the instrument s instrument number and the suffix code of the option you want to add Procedure 1 Press UTILITY and then the Overview s...

Страница 110: ...ted Viewing the System Information 6 To verify that the option has been installed view the system information on the instrument overview screen For instructions on how to display the overview screen s...

Страница 111: ...Configuration menu Press UTILITY and then the System Configuration soft key The following menu appears Formats the internal storage Confirm that you want to format internal memory CAUTION If you forma...

Страница 112: ...ce Life Built in printer Under normal conditions of use the period it takes to use 360 rolls of printer paper part number B9988AE LCD backlight Under normal conditions of use approximately 50000 hours...

Страница 113: ...of setting 20 mV Frequency bandwidth 3 dB 1 2 3 divp p sine wave input 1 M input measured from the probe tip when using the supplied 10 1 passive probe 10 1 conversion DLM5034 DLM5038 DLM5054 DLM5058...

Страница 114: ...eses are for high resolution mode Repetitive sampling mode 250 GS s Interpolation sampling mode 250 GS s Maximum record length Maximum record length for which repetitive acquisitions are possible No o...

Страница 115: ...e rejection is on 250 mV Minimum pulse width 5 ns 2 ns Maximum sample rate Real time sampling mode 1 25 GS s4 1 25 GS s Values inside parentheses are for high resolution mode 3 Repetitive sampling mod...

Страница 116: ...elected for CH1 to CH81 separately Noise rejection cannot be specified on channels set to TV trigger Low Approx 0 3 div of hysteresis Mid Approx 0 5 div of hysteresis High Approx 1 0 div of hysteresis...

Страница 117: ...and the condition changes to not met Time1 2 ns to 10 s 2 ns resolution 2 ns Time2 4 ns to 10 s resolution 2 ns Minimum spacing between Time1 and Time2 2 ns Outside Triggers when the time length duri...

Страница 118: ...bus signal The source can be selected from CH1 to CH8 1 Mode SOF Error ID Data ID OR Bit Rate 33 3k 83 3k 125k 250k 500k 1Mbps or User Define For User Define you can set a value from 10 k to 1 Mbps i...

Страница 119: ...720 60p 1080 25p 1080 24p 1080 24sF 1080 60p UserdefTV You can trigger on any TV signal by selecting standard or high definition setting the Hsync period and setting the sync guard Sync guard can be s...

Страница 120: ...erating conditions see section 6 12 after warm up 2 CH1 to CH4 for the DLM5034 DLM5054 6 4 Display Item Specifications Display 12 1 inch 30 7 cm color TFT LCD capacitive touch panel Display screen siz...

Страница 121: ...ned on and off collectively for CH1 to CH8 1 OFF Does not apply the specified offset to the result of cursor measurements computations and automated measurement of waveform parameters ON Applies the s...

Страница 122: ...al Input Section RealTime Achieves the required time scale by reducing the display record length Interpolation Performs interpolation sampling If you decrease the time scale further and the upper limi...

Страница 123: ...dow to 20 or 50 Display interpolation Select dot display of sample points sine interpolation display linear interpolation display or pulse interpolation display Graticule Dot grid line grid frame and...

Страница 124: ...d The following types and sub types can be selected for the FFT Type LS RS PS PSD CS TF CH Sub type MAG LOGMAG PHASE REAL IMAG Reference waveforms Eight reference waveforms Ref1 to Ref8 can be display...

Страница 125: ...Freq Period AveFreq Duty Pulse Count Delay In cycle mode the following items are valid Max Min P P High Low Amplitude Rms Mean Sdev IntegTY IntegTY Over Over The maximum total number of items that ca...

Страница 126: ...utomatically are listed below P Turn On On Turn Off Total Wp Turn On On Turn Off Total In addition statistical processing can be applied to the above items For details on statistical processing see St...

Страница 127: ...or measurements or automatically measured results are displayed they are printed below the waveform area Wide Saves a screen capture whose time axis is magnified twice The menu is not saved If cursor...

Страница 128: ...nd message language can be specified The click sound can be turned on and off Time synchronization feature Synchronization source Supports IEEE1588 2008 PTP v2 slave only Supports PTP packets of Layer...

Страница 129: ...See section 6 8 USB for Peripherals 6 8 USB Ports for Peripherals Item Specifications Connector type USB type A receptacle Electrical and mechanical USB Rev 2 0 compliant Supported transfer modes LS...

Страница 130: ...GER OUT Item Specifications Connector type BNC Output level 3 3 V CMOS Output impedance Approx 50 Output logic negative logic and positive logic switchable Output delay time 600 ns or less Output hold...

Страница 131: ...8 701929 Side Panel Probe Power Terminal option Item Specifications Output terminals DLM5034 DLM5054 4 DLM5038 DLM5058 8 Output voltage 12 V up to 2 0 A in combination with the front panel probe power...

Страница 132: ...as a device that conforms to one of the following two protocols USBTMC USB488 USB Test and Measurement Class Ver 1 0 Communication commands can be used through USB Mass Storage Class Ver 1 1 Only rea...

Страница 133: ...asurement start stop sampling clock clock time trigger Inter unit sampling skew When the connection cable 701982 01 is used 20 20 ns typical value When the connection cable 701982 02 is used 27 90 ns...

Страница 134: ...m battery Battery life Approx five years at an ambient temperature of 25 C Safety standards Compliant standards EN 61010 1 Overvoltage Category II1 Pollution degree 22 EN 61010 2 030 Measurement Categ...

Страница 135: ...d of a BNC cable Use cables that are 3 m or less in length Video signal output cable Attach a ferrite core TDK ZCAT2035 0930A YOKOGAWA A1190MN to each end of a D Sub 15 Pin VGA shielded cable Use cabl...

Страница 136: ...0 GHz to 6 0 GHz 3 V m Criteria A 3 Conducted immunity EN 61000 4 6 10 V Criteria A 4 EFT Burst EN 61000 4 4 Power line 2 kV Signal line 1 kV Criteria B 5 Surge immunity EN 61000 4 5 1 kV between line...

Страница 137: ...058 03EN Specifications 1 2 3 4 5 6 App Index 6 13 External Dimensions Instrument Unit mm Unless otherwise specified tolerances are 3 however tolerances are 0 3 mm when below 10 mm Rear view 426 9 266...

Страница 138: ...own on the screen changes to 1 25 GS s On the table under When the Set Record Length Is 125 kpoint provided later you can confirm that the sample rate is 1 25 GS s when the time scale Time div is set...

Страница 139: ...k 5 ms 25 k 1 25 k 25 k 1 25 k 25 k 1 25 k 25 k 1 25 k 25 k 1 25 k 25 k 1 25 k 2 ms 62 5 k 1 25 k 62 5 k 1 25 k 62 5 k 1 25 k 62 5 k 1 25 k 62 5 k 1 25 k 62 5 k 1 25 k 1 ms 125 k 1 25 k 125 k 1 25 k...

Страница 140: ...2 5 k 125 k 12 5 k 125 k 12 5 k 125 k 12 5 k 125 k 12 5 k 125 k 12 5 k 5 ms 250 k 12 5 k 250 k 12 5 k 250 k 12 5 k 250 k 12 5 k 250 k 12 5 k 250 k 12 5 k 2 ms 625 k 12 5 k 625 k 12 5 k 625 k 12 5 k 62...

Страница 141: ...25 M 125 k 1 25 M 125 k 5 ms 2 5 M 125 k 2 5 M 125 k 2 5 M 125 k 2 5 M 125 k 2 5 M 125 k 2 5 M 125 k 2 ms 6 25 M 125 k 6 25 M 125 k 6 25 M 125 k 6 25 M 125 k 6 25 M 125 k 6 25 M 125 k 1 ms 12 5 M 125...

Страница 142: ...ms 12 5 M 1 25 M 12 5 M 1 25 M 12 5 M 1 25 M 12 5 M 1 25 M 12 5 M 1 25 M 12 5 M 1 25 M 5 ms 25 M 1 25 M 25 M 1 25 M 25 M 1 25 M 25 M 1 25 M 25 M 1 25 M 25 M 1 25 M 2 ms 62 5 M 1 25 M 62 5 M 1 25 M 62...

Страница 143: ...5 M 2 5 M 2 5 M 50 ms 5 M 2 5 M 5 M 2 5 M 5 M 2 5 M 5 M 2 5 M 20 ms 12 5 M 2 5 M 12 5 M 2 5 M 12 5 M 2 5 M 12 5 M 2 5 M 10 ms 25 M 2 5 M 25 M 2 5 M 25 M 2 5 M 25 M 2 5 M 5 ms 50 M 2 5 M 50 M 2 5 M 50...

Страница 144: ...M 5 M 50 ms 10 M 5 M 10 M 5 M 10 M 5 M 10 M 5 M 20 ms 25 M 5 M 25 M 5 M 25 M 5 M 25 M 5 M 10 ms 50 M 5 M 50 M 5 M 50 M 5 M 50 M 5 M 5 ms 100 M 5 M 100 M 5 M 100 M 5 M 100 M 5 M 2 ms 250 M 5 M 250 M 5...

Страница 145: ...5 M 12 5 M 12 5 M 12 5 M 12 5 M 50 ms 25 M 12 5 M 25 M 12 5 M 25 M 12 5 M 25 M 12 5 M 20 ms 62 5 M 12 5 M 62 5 M 12 5 M 62 5 M 12 5 M 62 5 M 12 5 M 10 ms 125 M 12 5 M 125 M 12 5 M 125 M 12 5 M 125 M...

Страница 146: ...M 25 M 50 M 25 M 20 ms 125 M 25 M 125 M 25 M 125 M 25 M 125 M 25 M 10 ms 250 M 25 M 250 M 25 M 250 M 25 M 250 M 25 M 5 ms 500 M 25 M 500 M 25 M 500 M 25 M 500 M 25 M 2 ms 1 25 G 25 M 1 25 G 25 M 1 25...

Страница 147: ...M 50 M 500 M 50 M 500 M 50 M 5 ms 625 M 31 25 M 625 M 31 25 M 625 M 31 25 M 625 M 31 25 M 2 ms 2 5 G 50 M 2 5 G 50 M 1 25 G 25 M 2 5 G 50 M 1 ms 2 5 G 25 M 5 G 50 M 1 25 G 12 5 M 5 G 50 M 500 s 2 5 G...

Страница 148: ...s 625 M 125 M 625 M 125 M 625 M 125 M 625 M 125 M 10 ms 1 25 G 125 M 1 25 G 125 M 1 25 G 125 M 1 25 G 125 M 5 ms 2 5 G 125 M 2 5 G 125 M 1 25 G 62 5 M 2 5 G 125 M 2 ms 2 5 G 50 M 5 G 100 M 1 25 G 25 M...

Страница 149: ...5 M 250 M 5 s 5 M 250 M 5 M 250 M 2 s 12 5 M 250 M 12 5 M 250 M 1 s 25 M 250 M 25 M 250 M 500 ms 50 M 250 M 50 M 250 M 200 ms 125 M 250 M 125 M 250 M 100 ms 250 M 250 M 250 M 250 M 50 ms 500 M 250 M 5...

Страница 150: ...M 500 M 5 M 500 M 5 s 10 M 500 M 10 M 500 M 2 s 25 M 500 M 25 M 500 M 1 s 50 M 500 M 50 M 500 M 500 ms 100 M 500 M 100 M 500 M 200 ms 250 M 500 M 250 M 500 M 100 ms 500 M 500 M 500 M 500 M 50 ms 625 M...

Страница 151: ...Set the vertical position to 0 div CTRL V EDGE CTRL W ZOOM2 CTRL X B TRIG CTRL Y ZOOM1 CTRL Z CH1 CTRL 1 CH2 CTRL2 CH3 CTRL3 CH4 CTRL4 CH5 CTRL5 CH6 CTRL6 CH7 CTRL7 CH8 CTRL8 Executing an HELP CTRL T...

Страница 152: ...knob CTRL HOME Increase the trigger position POSITION knob CTRL PAGE UP Decrease the magnification ZOOM knob CTRL DELETE Decrease the vertical position VERTICAL POSITION knob CTRL END Decrease the tri...

Страница 153: ...Acquisition memory Display memory CPU Main memory LCD Touch panel Built in printer option USB for peripheral devices USB For connecting to a PC Ethernet GP IB option Internal storage Keyboard Video s...

Страница 154: ...At the A D converters the voltage levels are converted into digital data The digital data undergoes digital filtering as well as sampling and envelope processing at the sample rate corresponding to t...

Страница 155: ...9 Numerics Page 366973 cable 4 7 701937 probe 2 14 701949 probe 2 14 701988 probe 2 21 701989 probe 2 21 A Page accessories sold separately vii ACQUIRE key 1 7 ACTION GO NO GO key 1 6 additional opti...

Страница 156: ...st 5 9 SERIAL BUS key 1 8 Serial No 5 12 SET key 1 10 3 2 settings after the execution of auto setup 3 24 SHIFT key 1 10 signals that auto setup can be applied to 3 23 SINGLE key 1 7 SNAP SHOT 1 9 sta...