2-50
IM DL850E-01EN
Trace Settings (Trace Setup)
Display Groups (Select Display Gr.)
Sub Channels 1 to 60 (Sub Channel 1 to 60)
Configure the display label, zoom magnification, scaling method, and display range settings of the specified sub
channel.
Labels (Label)
Zooming by Setting a Magnification (V Zoom)
Configure this setting when Value Type is set to Logic.
Selectable range: x0.1 to x3.33
Scaling (Scale)
Configure this setting when Value Type is not set to Logic.
Auto: The upper and lower display range limits are set automatically.
Default: To the greatest extent possible within the settable range, the maximum and minimum values are used to
set the display range setting (which is explained below) of the specified sub channel.
Display Range (Upper/Lower)
Set the upper and lower limits of the waveform display range.
Selectable range: −30.000E+30 to +30.000E+30
Configuring the Scales of All Sub Channels (All SubChannel Scale)
Configure the scales of all sub channels.
All SubChannel Auto: The upper and lower display range limits are set automatically.
All SubChannel Default: To the greatest extent possible within the settable range, the maximum and minimum
values are used to set the upper and lower limits of all sub channels.
LIN Frame Format
Recessive
Dominant
Break
Field
Synch
Field
Identifier
Field
Data
Field
Checksum
8N (1 ≤ N ≤ 8)
LIN Frame
Header
Response
2 Vertical Axis