9-3
IM DL850E-01EN
Measurement Items (Item Setup)
You can measure the following horizontal values at the cursor positions.
X1
Time value at Cursor1
X2
Time value at Cursor2
ΔX
Difference between the time values of Cursor1 and Cursor2
1/ΔX
Reciprocal of the difference between the time values of Cursor1 and Cursor2
Y1
Vertical value at the intersection of Cursor1 and the waveform
1
Y2
Vertical value at the intersection of Cursor2 and the waveform
2
ΔY
Difference between the vertical values at the intersections of the waveform with
Cursor1 and Cursor2
2
1 When Trace is set to All, the values for the channels of all installed modules, the sub channels, and the
Math channel are measured.
2 This option does not appear when Trace is set to All.
Example of Logic Signal Measurement
When you measure logic waveforms using vertical cursors, the measured values of Y1 and Y2 are determined in
the manner shown below.
1
Cursor 1
Cursor 2
2
3
4
5
6
7
8
• When the cursor order is 1->8
Bit Y1: 01001010
Y2: 10110010
Hex Y1: 4A
Y2: B2
• When the cursor order is 8->1
Bit Y1: 01010010
Y2: 01001101
Hex Y1: 52
Y2: 4D
When No Bits Are Turned Off
Cursor 1
Cursor 2
1
2
4
5
7
8
• When the cursor order is 1->8
Bit Y1: 01-01-10
Y2: 10-10-10
Hex Y1: 16
Y2: 2A
• When the cursor order is 8->1
Bit Y1: 01-10-10
Y2: 01-01-01
Hex Y1: 1A
Y2: 15
When Some Bits Are Turned Off
In the logic settings in the preferences, you can set the display format (Bit or Hex), the cursor order (Cursor
Order), and the bit order (Bit Order).
9 Cursor Measurement