11-2
IM DL850E-01EN
Basic Arithmetic (S1+S2, S1
−
S2, S1*S2, and S1/S2)
Performs addition, subtraction, multiplication, or division on the two waveforms assigned to Source1 and
Source2.
+ computed
waveform
– computed
waveform
× computed
waveform
Computation source
waveforms
Computed waveform
Computation Source Waveforms (Source1 and Source2)
CH1 to CH16,
1
16chVOLT,
2
16chTEMP/VOLT,
2
CAN,
3
LIN,
3
Math1 to Math7
4
1 You can select the channel of an installed module. However, you cannot select the channel of a logic module.
2 When a 16-CH voltage input module or 16-CH temperature/voltage input module is installed. After you select
16chVOLT or 16chTEMP/VOLT, select a sub channel.
3 On the DL850EV when a CAN bus monitor module or CAN & LIN bus monitor module is installed. After you
select CAN or LIN, select a sub channel. You cannot select waveforms whose data type (Value Type) is set to
Logic.
4 You can use other computed waveforms as computation source waveforms. If the waveform that you are
configuring is MathX, you can use a computation waveform up to MathX − 1 as the computation source
waveform. You cannot use another computation waveform as the computation source waveform for Math1.
When computation is performed on a linearly scaled channel, the scaled values are used.
Unit (Unit)
You can assign a unit of up to four characters in length to the computed results. The specified unit is reflected in
the scale values.
Label (Label)
You can create a label of up to eight characters in length. The labels that you create are displayed on the screen.
Turning the Display On and Off (Display)
Select whether to display the computed waveform.
• ON: Displays the computed waveform
• OFF: Does not display the computed waveform
Binary Conversion (Bin (S1))
Using the specified threshold levels, you can convert the waveform assigned to Source to a digital waveform.
1
0
Upper
Lower
Computation Source Waveform (Source), Unit (Unit), Label (Label), and Turning the Display On and Off
(Display)
These settings are the same as those for basic arithmetic.
Upper and Lower Thresholds (Thr. Upper/Thr. Lower)
Set the upper and lower threshold values. All values above the upper threshold on the computation source
waveform are converted to ones, and all values below the threshold are converted to zeros.
11 Computation