13 Messages and Corrective Actions
This section contains only the status and error messages that have been added for the
Power Analysis Function (/G4 Option). For a description of the standard messages see
section 16.2 in the
DL7440/DL7480 User’s Manual IM701450-01E
or appendix 2 in the
DL7440/DL7480 Communication Interface User’s Manual IM701450-17E
(CD-ROM).
Code
Message
Corrective Action
Reference Section
43
Aborted the auto deskew processing.
–
Chapter 4
44
Auto deskew cannot be executed in
–
Chapter 4, *
the following cases.
• When the trigger type is not Simple
• When the trigger type is Simple and the source is Ext or Line
• When Trigger Source = Deskew Target CH
• When Deskew Target CH is Pod A or Pod B
45
Auto deskew is in progress.
Wait until auto deskew completes.
Chapter 4
46
Cycle measure is not executed.
Execute Cycle measure of the trend target item before
Chapter 8
displaying the trend waveform.
873
Invalid math operation.
Check the math operation that you are trying to specify.
Chapter 7
874
Invalid math source.
Check the math source that you are trying to specify.
Chapter 7
875
Invalid measure item.
Check the measurement item that you are trying to
Chapter 5,
specify.
Chapter 7 to 10
*
See the
DL7440/DL7480 User’s Manual IM701450-01E
.