3-3
IM 701510-01E
Before Starting Observation and Measurement of W
aveforms
3
3.3
Connecting the SCSI cable (for DL1520L with
suffix code -C4)
Name, position and pin cofigurationo of the SCSI cable connector
• Name
: OPTION
• Position : Real panel of DL1520L
• Pin configuration
Pin No.
Signal name
Pin No. Signal name
Pin No.
Signal name
1 to 12
GND
32
-DB6
43
-BSY
13
NC
33
-DB7
44
-ACK
14 to 25
GND
34
-DBP
45
-RST
26
-DB0
35 to 37
GND
46
-MSG
27
-DB1
38
TERMPWR
47
-SEL
28
-DB2
39,40
GND
48
-C/D
29
-DB3
41
-ATN
49
-REQ
30
-DB4
42
GND
50
-I/O
31
-DB5
1
25
26
50
2
27
3
28
24
49
23
48
(Rear panel)
Pin No.
OPTION
Other necessary equipment
SCSI Device
SCSI devices which can send and receive data over SCSI and can set the ID number to "5"
can be connected the instrument. There are exceptions as in the case when the media used
on the SCSI device is formatted with another equipment such as a PC, or the pin
configuration of the connector is different. Refer to section 16.7, for the SCSI specification.
Also, for information on general precautions in using the SCSI device, refer to the
instruction manual for that device.
SCSI cable
SCSI cables should have the connector shape and pin configuration which match the
"OPTION" connector on the rear panel of the DL1520L (refer to the figure above). They
should be less than 3 m long with a characteristic impedance between 90 and 132
Ω
. Also,
attach a ferrite core (TDK:ZCAT2035-0930A) on each end of the cable.
Terminator
Some SCSI devices require terminators. Follow the instruction manual for the SCSI device.
Note
DL1540/DL1540L with software (ROM) version 1.20 or later can save/load data from the SCSI device
through the SCSI interface unit 700930 (sold separately). For details, refer to the instruction manual for