IM 701510-01E
15-5
T
roubleshooting, Maintenance
and Inspection
15
No.
Message
Reference Page
804
(For DL1520)
It is not possible to set SINGLE as the trigger mode while averaging mode is active.
(For DL1540)
It is not possible to set SGL(S)/SGL(L)/N-SGL as the trigger mode or history mode while averaging mode
is active.
(For DL1520L/DL1540L)
It is not possible to set SINGLE/N-SGL as the trigger mode or history mode while averaging mode is active. 6-13, 7-9
805
(For DL1520)
It is not possible to set averaging mode while SINGLE as the trigger mode.
(For DL1540)
It is not possible to set averaging mode while SGL(S)/SGL(L)/N-SGL as the trigger mode or history mode
is active.
(For DL1520L/DL1540L)
It is not possible to set averaging mode while SINGLE/N-SGL as the trigger mode or history mode is active. 7-2
806
It is not possible to turn ON the smoothing function in Envelope mode.
7-3
807
It is not possible to set Envelope mode while smoothing function is turned ON.
7-2
808
It is not possible to edit a zone waveform during dual display, X-Y waveform display, FFT waveform
display or during loading/recalling a waveform.
9-6, 7
809
The operation is not possible in GO/NO-GO zone type.
—
810
The trace to be edited is not currently displayed.
9-6
811
Zone data does not exist.
9-2, 7
812
It is not possible to set FFT mode in Envelope mode.
8-18
813
It is not possible to set Envelope mode in FFT mode.
7-2
814
It is not possible to set FFT mode within the repetitive sampling range.
8-18
815
It is not possible to set the repetitive sampling range in FFT mode.
7-1
816
(For DL1540)
It is not possible to set history mode in SGL(S)/SGL(L)/N-SGL as the trigger mode.
(For DL1520L/DL1540L)
It is not possible to set history mode in SINGLE/N-SGL as the trigger mode.
7-9
817
(For DL1540)
It is not possible to set SGL(S)/SGL(L)/N-SGL as the trigger mode in history mode.
(For DL1520L/DL1540L)
It is not possible to set SINGLE/N-SGL as the trigger mode in history mode.
7-8
818
It is not possible to change a GP-IB address when talk-only mode is active.
*
819
Set the correct date and time.
3-9
820
No keys can be operated during real time printing except the START/STOP and COPY key.
10-10
821
It is not possible to set Envelope mode during an external clock.
7-2
822
(For DL1520)
It is not possible to set SINGLE as the trigger mode in the time range.
(For DL1520L)
At this time range setting, it is not possible to set N-SGL trigger mode, the history mode or the record length
to 400K/1M words.
(For DL1540)
It is not possible to set SGL(L)/N-SGL as the trigger mode or history mode in the time range.
(For DL1540L)
At this time range setting, It is not possible to set N-SGL trigger mode, the history mode or the record length
to 400K/1M/2M words.
5-11,6-13,7-10
823
The trace to be operated is not currently loaded/recalled.
—
824
It is not possible to set H-position during roll mode acquisition.
5-12
825
The operation is not possible in acquisition hold.
12-6
826
(For DL1540/DL1540L)
The operation is not possible while Isolator is not connected.
—
827
(For DL1540/DL1540L)
The operation is not possible while Isolator is connected.
—
* Refer to the Communication Interface User’s Manual (IM701510-11E).
15.2
Error Messages and Corrective Actions