6-2
IM 04L42B01-01E
• Recording Conditions of Event Data
Press
MENU
(to switch to setting mode), and select the
Menu
tab >
Data save
>
Event data
.
Setup Items
• Memory >
Data kind
Settings
Description
Display
Records display data.
E+D
Records display data and event data. This setting cannot be selected
when the trend interval switching function is set to On. This setting cannot
be selected when the multi batch function (/BT2 option) is being used or
on DXs with the /AS1 advanced security option.
Event
Records event data.
• Memory sample > On/Off
Turn On the target channels.
• Trend/Save interval > Trend interval [/div] (when recording display data)
See the table below. You can only set trend intervals that are longer than the scan
interval you set in Basic Setting Mode.
• Trend/Save interval > Save interval (when recording display data)
Select the size of a record data file. The recorded data is divided by the file size
specified here. The available settings vary depending on the number of memory
sampling channels and the
Trend interval
setting.
Trend interval
*1
5 s
*2
10 s
*2
15 s
*3
30 s
1 min
Sample rate
125 ms
250 ms
500 ms
1 s
2 s
Selectable range
of auto save
interval
10 min to 12
hours
10 min to 1
day
10 min to 3
days
10 min to 7
days
10 min to 14
days
Trend interval
*1
2 min
5 min
10 min
15 min
20 min
Sample rate
4 s
10 s
20 s
30 s
40 s
Selectable range
of auto save
interval
10 min to 14
days
10 min to 31
days
10 min to 31
days
10 min to 31
days
1 hour to 31
days
Trend interval
*1
30 min
1 h
2 h
4 h
10 h
Sample rate
1 min
2 min
4 min
8 min
20 min
Selectable range
of auto save
interval
1 hour to 31
days
1 hour to 31
days
2 hours to
31 days
4 hours to
31 days
8 hours to
31 days
*1 You cannot set a trend interval that corresponds to a sampling interval that is faster than the
scan interval.
*2 Selectable on the DX2004 and DX2008 (release number 3 or later).
*3 Selectable in fast sampling mode on the DX2010, DX2020, DX2030, DX2040, and DX2048
(release number 3 or later).
• Trend/Save interval > Second interval [/div]
See section 5.3.
6.1 Setting the Recording Conditions of the Measured Data
Содержание Daqstation DX2000
Страница 2: ......
Страница 98: ...Blank...
Страница 132: ...Blank...
Страница 224: ...Blank...
Страница 292: ...Blank...
Страница 324: ...Blank...
Страница 348: ...13 24 IM 04L42B01 01E 13 7 External Dimensions See the DX2000 Operation Guide IM04L42B01 02E...
Страница 366: ...Blank...