8-12
IM CA150E
SD
Normal Setting
Sets and queries the source values
○
×
Command = SDm<Delimiter>
→
Answer = SDm <Delimiter>
Command = SD?<Delimiter>
→
Return = SDm <Delimiter>
Parameter
m = Source Values
100m V
m = 0 to ±110.000 mV
1 V
m = 0 to ±1.10000 V
10 V
m = 0 to ±11.0000 V
30 V
m = 0 to ±30.00 V
20 mA, 4-20 mA
m = 0 to +22.000 mA
20 mA SINK
m = 0 to -22.000 mA
500 Ω
m = 0 to 550.00 Ω
5 kΩ
m = 0 to 5.5000 k
Ω
50 kΩ
m = 0 to 55.000 k
Ω
PT100
m = -200.0 to 850.0 °C
PT100 (ITPS-68)
m = -200.0 to 650.0 °C
JPT100
m = -200.0 to 500.0 °C
K
m = -200.0 to 1372.0 °C
E
m = -200.0 to 1000.0 °C
J
m = -200.0 to 1200.0 °C
T
m = -200.0 to 400.0 °C
N
m = -200.0 to 1300.0 °C
L
m = -200.0 to 900.0 °C
U
m = -200.0 to 400.0 °C
R
m = 0 to 1768 °C
S
m = 0 to 1768 °C
B
m = 600 to 1820 °C
100 Hz (*1)
m = 1.00 to 110.00 Hz
1000 Hz (*1)
m = 90.0 to 1100.0 Hz
10 kHz (*1)
m = 0.9 kHz to 11.0 kHz
50 kHz (*1)
m = 9 kHz to 50 kHz
1000 CPM
m = 1.0 to 1100.0 CPM
PULSE DCV (*2)
m = 0 to 11.0000 V
PULSE Cycle (*3) m = 0 (cont), 1 to 60000 cycle
*1: When the PULSE source display setting is frequency, only (PU0) can
be set
*2: When the PULSE source display setting is amplitude, only (PU1) can
be set
*3: When the PULSE source display setting is pulse number, only (PU2)
can be set
SF
Normal Setting
Sets and queries the source function
○
×
Command = SFm<Delimiter>
→
Answer = SFm <Delimiter>
Command = SF?<Delimiter>
→
Return = SFm <Delimiter>
Parameter
m = 0: DCV/1: DCA/2: Ω
/3: TC/4: RTD/5: Pulse
Set the PULSE source DCV and cycle of each parameter with the PU
command.
8.3 Detailed Description of Commands