11-2
IM CA150E
TC output
Pulse
output
R
S
B
0 to 1768 °C
0 to 1768 °C
600 to 1820 °C
±(0.02% +1.2 °C)
However, 0 to 100 °C
±(0.02% +2 °C)
±(0.02% +1.2 °C)
However, 0 to 100 °C
±(0.02% +2 °C)
±(0.02% +1 °C)
However, 600 to 1000 °C
±(0.02% +1.5 °C)
CPM
1.0 to 1100.0 CPM ±0.5 CPM
0.1 CPM
100 Hz
1.00 to 110.00 Hz
±0.05 Hz
0.01 Hz
1000 Hz
90.0 to 1100.0 Hz
±0.5 Hz
0.1 Hz
10 kHz
0.9 kHz to 11.0 kHz ±0.1 kHz
0.1 kHz
50 kHz
9 kHz to 50 kHz
±1 kHz
1 kHz
Output voltage
(+0.1 V to +11 V) ±10%
(zero base waveform)
Maximum load current:
10 mA
Pulse number
Continuous, 1 to 60,000
cycles
Range
Source Range
Accuracy
Remarks
Setting
Resolution
1 °C
*3 TC source accuracy
does not include RJ
sensor accuracy.
RJ Sensor Specification
Measurement range:
-10 to 50 °C
Accuracy
(when combined with
main unit)
18 to 28 °C: ±0.5 °C
Other than 18 to 28 °C:
±1 °C
*1
*3
Temperature coefficient: When in the ranges of 0≤T<18°C and 28<T≤40°C,
add the accuracy shown above ×(1/10)/°C.
The above accuracy ranges are for values obtained
when the instrument has been left for two hours after charging ends (or is stopped).
*1: ITS-90 or IPTS-68 can be set with the internal settings.
*2: Excitation current Is: When less than 0.1 mA to 1 mA,
then add {0.05/ls (mA)} (Ω) or {0.12/ls (mA)} (°C).
About Accuracy during Charging
Accuracy is affected by the heat generated during charging or immediately
after charging ends (is stopped).
Refer to the accuracy values shown in "Source during Charging" on the next page.
11. Specifications