Source
4-9
IM CA150E
1
2
3
4
5
6
7
8
9
10
11
12
App
1
Connect the terminals.
2
Use the FUNCTION key on the SOURCE side to align
the
source mark with Ω. (Resistance selection)
3
Use the RANGE key to select a range. (500 Ω, 5 kΩ, or 50 kΩ)
4
Use the ▲ ▼ output value setting keys to set each digit output value.
Each ▲ ▼ key corresponds to a digit of the setting value.
Each time a key is pressed the value is incremented or decremented by
one count. If you attempt to increment or decrement
the value 9 or 0, the digit moves up or down one place.
A value continues to change when you hold down the key.
Press the CLEAR key to restore the setting to its initial value (zero).
5
Press the SOURCE ON/OFF key to start generation.
(SOURCE OFF changes to ON.)
6
Press the SOURCE ON/OFF key again to turn off the output.
(OFF lights and the output terminals are open-circuited.)
• When using the divided output (n/m) function,
refer to "4.8 Divided Output (n/m) Function."
• When using a sweep output function,
refer to "4.9 Sweep Output Functions."
• When using the temperature monitor function,
refer to "4.10 Temperature Monitor Function."
Tip
If settings are modified with the FUNCTION and RANGE keys,
the protection circuit is activated and the output is turned off.
4.4 Source Resistance (Ω) Signals