19
IM AQ6373B-02EN
You can also shut down the instrument using panel keys and soft keys.
1.
Press
SYSTEM
.
2.
Press the
MORE
soft key three times.
The SYSTEM
4/4 screen is displayed.
3.
Press the
SHUT DOWN
soft key.
4.
Press the
YES
soft key. Shut down begins.
5.
After the
POWER
switch changes from green to orange, turn OFF the
MAIN
POWER
switch on the rear panel of the instrument.
Note
If for some reason the instrument fails to shut down normally, hold down the
POWER
switch
for approximately four seconds or longer to force standby mode. Note that the operating system
configuration file will not be backed up, possibly resulting in malfunctions upon start up the next
time the instrument is turned on.
Explanation
Screen when the instrument was not shut down
If the shutdown procedure was not performed after the previous session, the following
message appears after start up.
If you do not perform a shutdown procedure, the monochromator will not be retracted.
Transporting or moving the instrument repeatedly in this condition may damage the
monochromator.
When turning OFF the power, always perform the shut down procedure.
Press any key to clear this message.
7 Turning the Power ON/OFF