![YOKOGAWA AQ6370D Скачать руководство пользователя страница 43](http://html.mh-extra.com/html/yokogawa/aq6370d/aq6370d_getting-started-manual_911211043.webp)
27
IM AQ6370D-02EN
Item
Specifications
Electrical interfaces
GP-IB, RS-232, Ethernet, USB, SVGA output, analog output port, trigger input port, trigger
output port
Remote control
12
GP-IB, RS-232, Ethernet (TCP/IP)
AQ6317 series compliant commands (IEEE488.1) and IEEE488.2
Data storage
Internal storage
512 MB or higher
External storage
USB storage media (USB memory/HDD), format: FAT32
File types
CSV (text), binary, bitmap, TIFF
Display
13
10.4" color LCD (resolution: 800 x 600 pixels)
External dimensions
Approximately 426 (W) x 221 (H) x 459 (D) mm (excluding the protector and handle)
Mass
Approximately 19 kg
Power
requirement
100 to 240 VAC, 50/60 Hz, approximately 100 VA
Environment conditions
Optimal temperature: +18 to +28°C
Operating temperature range: +5 to +35°C
Storage temperature range: -10 to +50°C
Ambient humidity: 20 to 80% RH or less (no condensation)
Recommended calibration period 1 year
Function
Measurement
Setting of measuring conditions Center wavelength, span, wavelength sampling points, wavelength resolution, measurement
sensitivity, high dynamic mode, average count (1 to 999), double-speed measurement mode,
smoothing, APC level correction
Sweep settings
Single sweep, repeat sweep, AUTO (automatically sets measuring conditions), sweep
between marker, data logging
Measurement function
CW measurement, pulse light measurement, external trigger measurement, gate
measurement, air/vacuum wavelength measurement
Other
Sweep status output, analog output
Display
Vertical scale
Level scale (0.1 to 10 dB/div, linear), level subscale (0.1 to 10 dB/div, linear), reference level
display, DIV display (8, 10, 12), percentage display, dB/km display, power density (dB/nm)
display, noise mask
Horizontal scale
Horizontal wavelength (nm) display, frequency (THz) display, zoom in/zoom out display
Display mode and items Single waveform display, split screen display, data table display, label display, template
display, measurement condition display
Trace
Display function
Simultaneous display of 7 independent traces, max/min value detection display, calculation
between traces display, normalized display, roll averaging (sweep average) display (2 to 100
times), curve fit display, peak curve fit display, marker curve fit display
Other
Trace copy, trace clear, write mode fixed mode setting, show/hide setting
Marker and search
Marker
Delta marker (1024 points max.), vertical/horizontal line markers, advanced marker
Search
Peak search, bottom search, auto search (ON/OFF), search between vertical axis line
markers, search within zoom area
Data analysis
Analysis feature
Spectral width analysis (threshold, envelope, RMS, Peak-RMS, notch), WDM (OSNR)
analysis, EDFA-NF analysis, filter peak/bottom analysis, WDM filter peak/bottom analysis,
DFB-LD analysis, FP-LD analysis, LED analysis, SMSR analysis, power analysis, PMD
analysis, Pass/Fail judgment from template
Other
Auto analysis execution setting, analysis between vertical axis line markers, analysis within
the zoom area
Auto measurement
Programming
64 programs, 200 steps/program
Other
Alignment
Auto alignment using built-in calibration light source (when the built-in light source
specification is -L1) or an external light source
Wavelength Calibration Auto wavelength calibration using built-in calibration light source (when the built-in light
source specification is -L1) or an external light source
Resolution calibration
User-defined resolution calibration using an external light source (DFB-LD)
11 Specifications