48
IM AQ1210-02EN
Performing Averaged Measurements
4.1, “Performing Averaged Measurements in TRACE Mode” in the User’s Manual
In averaged measurements, the data that is acquired from each pulse is averaged and displayed.
1.
Tap
OTDR
on the MENU screen.
2.
Tap
Wavelength
to select the wavelength to be measured.
3.
Press
AVG
to start a measurement. If you press AVG again before measurement completion, the measurement stops at that point.
Data display
screen
Displays the
averaged
waveform
Laser on indication
Averaged measurement
in progress indication
Measurement ends when 100% is reached.
You can only control the markers and edit
labels during measurement. When the
measurement is finished, the instrument
returns to the previous soft key display.
Detects events from the acquired
waveform
You cannot set the
wavelength during
measurement.
Waveform Measurement