Setting the Analysis Conditions (Analysis)
2.2, “Analysis (Analysis) Conditions” in the User’s Manual
1.
Tap
OTDR
on the MENU screen.
2.
Press
SETUP
to display the setup screen.
3.
Tap the
ANALYSIS
tab to display the following screen.
ANALYSIS tab
Splice loss threshold
(0.01dB to 9.99dB)
An event is detected
when this value is
exceeded.
End of fiber threshold
(3dB to 65dB)
An event E is detected
when this value is
exceeded.
Bending loss threshold
(0.001dB to 99.999dB)
An event (bending) is
detected when this value
is exceeded.
Setting a Launch Fiber
Set the conditions for
when a launch fiber is
attached to the start or
end face.
Set the approximation
method (TPA, LSA).
Set the backscatter level.
Return loss threshold (20dB to
70dB).
An event is detected when this value
is exceeded.
Splitter detection threshold
(1dB to 20dB)
An event (splitter) is detected when
this value is exceeded.
Configure the pass/fail judgment.
Set a pass/fail judgment threshold
on each measurement item.
Set the event search
approximation method
(TPA, LSA).
Executes initialization
Press to reset the settings to
their factory defaults.
Setup