IM AQ1000-02EN
25
Charging the Instrument
• When the instrument’s battery is low, a warning message will appear.
• When the battery is low, connect the USB-AC adapter to an electrical outlet, and charge the instrument. For the estimated battery
level, check the battery level indicator at the top of the screen.
Battery power level
CHARGE LED (red)
Solid:
Charging
Off:
Charging complete
Blinking:
Charging suspended
Power switch (POWER LED)
Green:
Running
Red:
Battery low
CHARGE
Power switch
Sufficiently full
Half full
Almost empty
The power switch (POWER LED) lights in red,
a warning message appears on the screen.
Power supply status display
USB-AC adapter connected
Charging
Battery level indicator
• Charging takes about 5 hours with the power turned off. If the instrument is charged with the power turned on, it may take longer
than 15 hours, but charging will be suspended after about 15 hours by the protection circuit. If battery charging does not complete
within 8 hours with the power turned off, stop immediately. The battery in the instrument may be malfunctioning. Contact your nearest
YOKOGAWA dealer.
Note
Over Discharge and Long Periods of Storage
• If you do not use the instrument for an extended period of time, the battery in the instrument may become over discharged. This
shortens the service life of the battery. To prevent over discharging, charge the instrument once a month.
• Charge the instrument prior to its first use or if it has not been used for an extended period of time.
Making Preparations for Measurements