9-4
IM 704420-01E
Error in communication query
Code
Message
Description and Corrective Action
Section
410
Query INTERRUPTED
Query transmission was aborted. Check the order
8.6
of transmission and reception.
420
Query UNTERMINATED
There is no response that can be transmitted.
8.6
Check the order of transmission and reception.
430
Query DEADLOCKED
Deadlock occurred. Aborting transmission.
8.6
Set the length of a program message including the
<PMT> to less than or equal to 1024 bytes.
440
Query UNTERMINATED after
The order to request the response is not correct. —
indefinite response
Do not specify a query after the *IDN? or *OPT?
command.
Error in system operation
Code
Message
Description and Corrective Action
Section
906
Fan stopped. Turn OFF the
The cooling fan has stopped. Immediately turn
3.4
power immediately.
OFF the power.
909
No Battery.
The backup battery is dead. Servicing is required 3.4
for battery replacement.
910
Calibration data lost.
The calibration value is abnormal. Servicing is
—
required for calibration.
912
Fatal error in Communication
Communication driver error. Servicing required.
—
-driver
914
Time out occurs in
Communication timeout error. Make sure to receive —
Communication
the data within the timeout time. There may a
problem in the communication line.
Miscellaneous
Code
Message
Description and Corrective Action
Section
350
Queue overflow
Read the error queue. Occurs when there are 16
8.8
or more messages in the error buffer.
Note
Code “350” occurs when the error queue overflows. This error is output only during a
STATus:ERRor?
query and does not appear on the screen.
9.2 Error Code Description and Corrective Actions