8
IM 701949-01EN
How to Adjust the Probe Phase
CAUTION
Do not apply excessive force to the phase adjustment screw.
Doing so may damage the internal variable capacitor.
French
ATTENTION
N’appliquez pas une force excessive sur la vis de réglage de
phase. Cela pourrait endommager le condensateur variable
interne.
1.
Connect the BNC connector of the probe to the input of the
oscilloscope, and connect the probe input to the oscilloscope signal
output terminal for probe compensation adjustment (CAL or COMP
terminal).
2.
Operate the vertical and time scales on the oscilloscope, and
turn the phase adjustment screw with the included adjustment
screwdriver to adjust the observed waveform to the correct
waveform (square wave). Adjust the probe capacitance to match
the input capacitance of the oscilloscope with the variable capacitor
inside the probe.
Correct Waveform
Over Compensation
Inadequate Compensation
3. How to Use