8

IM 701946-01EN

1  The product is for measurement category II (CAT II). Do not use it with measurement category III 

(CAT III), nor measurement category IV (CAT IV). When using devices or accessories with different 

measurement categories, the lower measurement category applies. See below for definitions of 

measurement categories.

2  Pollution degree refers to the degree of adhesion of a solid, liquid, or gas which deteriorates 

withstand voltage or surface resistivity. Pollution Degree 2 applies to normal indoor atmospheres 

(usually with only non-conductive pollution).

3  For conformity to environmental regulations and/or standards other than EU, contact your nearest 

Yokogawa office (PIM 113-01Z2).

Definitions of Measurement Categories

Measurement Category “O” (Other)

It applies to measurement of a circuit that is not 

connected directly to the main power source.

Measurement Category II (CAT II

)

It applies to measurement of electrical equipment 

that is powered through a fixed installation such 

as a wall outlet wired to a distribution board and 

measurement on such wiring.

Measurement Category III (CAT III)

It applies to measurement at the distribution level, 

that is, building wiring, fixed installations.

Measurement Category IV (CAT IV)

It applies to measurement at the primary supply 

level, that is, overhead lines, cable systems.

Voltage Derating

  

CAUTION

As the frequency of the input signal increases, the maximum 

input voltage of the probe decreases. See Specifications for the 

appropriate input voltage.

French

  

ATTENTION

Lorsque la fréquence du signal d’entrée augmente, la tension 

d’entrée maximale de la sonde diminue. Voir les spécifications 

pour la tension d’entrée appropriée.

  

Frequency [Hz]

Am

pl

itu

de

 A

rm

[V] s

in

us

Typical Voltage Derating

0

100

200

300

400

100 k

1 M

10 M

100 M

1 G

Содержание 701946

Страница 1: ...User s Manual Model 701946 Miniature Passive Probe IM 701946 01EN 5th Edition ...

Страница 2: ... us to serve you best by completing the user registration form accessible from our website https tmi yokogawa com support Contact Us If you want to resolve a technical support issue or need to contact YOKOGAWA please fill out the inquiry form on our website https tmi yokogawa com contact PIM 103 06E A5 ...

Страница 3: ...contacts Notes The contents of this manual are subject to change without prior notice as a result of continuing improvements to the product s performance and functionality The figures given in this manual may differ from those that actually appear on your product Every effort has been made in the preparation of this manual to ensure the accuracy of its contents However should you have any question...

Страница 4: ...ries II defined in IEC 61010 031 If this probe is operated in a manner not specified in this manual this may cause the protection capability of this product to lessen Additionally YOKOGAWA shall not be held responsible for defects arising from negligence of such warning and caution and also shall not guarantee the product in such case Before using this probe thoroughly read the instruction manual ...

Страница 5: ...ximum input voltage to the probe When the oscilloscope s input coupling is AC a DC voltage is applied to the oscilloscope s input at the same electric potential as the probe s input Make sure not to exceed the oscilloscope s maximum input voltage French AVERTISSEMENT But du produit Le produit est utilisé en association avec un oscilloscope pour observer et mesurer des signaux électriques Ne l util...

Страница 6: ... d entrée maximale sur la sonde Lorsque le couplage d entrée de l oscilloscope est AC une tension DC est appliquée à l entrée de l oscilloscope au même potentiel électrique que l entrée de la sonde Assurez vous de ne pas dépasser la tension d entrée maximale de l oscilloscope Regulations and Sales in Various Countries and Regions Waste Electrical and Electronic Equipment WEEE EU WEEE Directive val...

Страница 7: ...ation on the next page No Standard Accessories No Standard Accessories 1 Probe main unit 11 Protection cap 2 Spring tip Ø 0 5 mm 12 Ground lead 3 Rigid tip Ø 0 5 mm 13 Ground spring 4 Pincher tip 14 Ground blade 5 IC cap 0 5 mm pitch green 15 Cu sheets 2 x 2 cm x 2 sheets 6 IC cap 0 65 mm pitch blue 16 PCB adapter kit 7 IC cap 0 8 mm pitch gray 17 Marker tips 4 colors x 3 8 IC cap 1 0 mm pitch bro...

Страница 8: ...UNDER TEST 2 Spring tip Ø0 5 mm 3 Rigid tip Ø0 5 mm 4 Pincher tip length 77 6 mm 6 0 65 mm pitch blue 7 0 8 mm pitch gray 8 1 0 mm pitch brown 9 1 27 mm pitch black 17 Marker tip 4 colors x 3 green blue red yellow 19 Two footer positioner 10 Insulation cap 20 Manuals 106 5 15 2 5 87 8 18 6 Ø15 0 Ø9 2 Probe tip Ø2 5 mm Probe cable Adjustment screw Contact tip Ø0 5 mm 1 Dimensions of probe main unit...

Страница 9: ...2 See the Compliant Standards shown below See also Voltage Derating Electrical Characteristics Input Resistance system 10 MΩ 1 Input Capacitance system 9 5 pF typical value Input Impedance system See also Input Impedance Compensation Range 10 pF to 25 pF typical value Mechanical Characteristics Weight probe only Approximately 48 g Cable Length Approximately 1 2 m Environmental Specifications Altit...

Страница 10: ...a circuit that is not connected directly to the main power source Measurement Category II CAT II It applies to measurement of electrical equipment that is powered through a fixed installation such as a wall outlet wired to a distribution board and measurement on such wiring Measurement Category III CAT III It applies to measurement at the distribution level that is building wiring fixed installati...

Страница 11: ...0 1 k 10 k 100 k 1 M 10 M 100 M 1 G Frequency Hz Z Ω Handling CAUTION Since the contact tips are very thin and sharp handle them with care Avoid shock to the probe body Do not bend or pull the cables excessively Doing so may damage or disconnect the probe French ATTENTION Les pointes de contact étant très fines et pointues manipulezles avec precaution Évitez les chocs sur le corps de la sonde Ne p...

Страница 12: ...und blade on the probe you can minimize the length of the ground connection By using the included IC caps you can easily make connections to the pins of fine pitch ICs There are five types of IC caps included These IC caps have a pitch ranging from 0 5 mm to 1 27 mm Ground blead Cu pad To Ground pin Installing the Pincher Tip To use the pincher tip first install it on the probe so that the pincher...

Страница 13: ...ng the measurement Changing the Contact Tip To change the contact tip grip the contact tip firmly with pliers and carefully pull it straight out of its contact socket along the axis of the probe Do not grip the white plastic insulator or the housing with pliers because this will damage the probe If the contact tip is removed the new contact tip can be inserted with pliers into the socket along the...

Отзывы: