www.yictechnologies.com
It is important to activate all of the probes under the DUT to have a complete view of the
emissions on the DUT.
Amplitude Adjustment Tab
You can use an LNA or attenuator with EMScanner to increase or decrease the power level
of the signals coming from the scanner and going to the spectrum analyzer. To activate, click
on Amplification Setting or Attenuator Setting and then select the value from the drop-down
menu or type the value. The displayed value of the signal will be corrected for the value of
the gain or attenuation.
Internal Analyzer Pre-Amp:
This feature provides extra sensitivity. It is ON by default. Turn
this feature OFF if your spectrum analyzer is not equipped with a Pre-Amp.
Analyzer Tracking Generation:
This feature will remotely turn on the tracking generator of the
spectrum analyzer. When you run Spectral or Spectral/Spatial Scan, the tracking generator
will output a frequency that tracks the swept frequency. The Tracking Generation Feature is
currently available for the Keysight FieldFox N9912A. New drivers must be developed on
demand for any other spectrum analyzer; contact Y.I.C. Technologies
(
)