Maintenance 

Filter 

Keyway

Pull straight out 

Unscrew tip to 

replace

 filter 

Push straight on (do not 

twist) to install sensor 

 

Batteries: 

Install Batteries

: Remove screw located at rear end of unit and 

flip up the hinged battery door to open as shown.  Always insert all 

four batteries into the battery compartment in the correct direction. 

Note  polarity  mark  on  the  inside  of  the  battery  door  for  proper

battery orientation (see illustration on page 4).  

Sensor: 

Filter Replacement

: Unscrew sensor tip as shown to replace filter.

Replace  filter  whenever  it  becomes  visibly  dirty  or  every  2  to  3

months depending on use. 

Note: Never clean dirty filters with a solvent or soap and water.

Always replace with a new filter supplied with the leak detector

or they can be re-ordered from your supplier or distributor. 

Sensor Failure Mode:

 If the sensor is not working correctly, the H

2

Leak  Detector  will  not  come  out  of  the  warm-up  mode.  If  this

happens  first  be  sure  the  sensor  is  plugged  in  all  the  way.  If  that

does not correct the situation, replace the sensor. 

Sensor  Replacement

:  Remove  sensor  by  pulling  out  of  socket.

Install the new sensor by aligning the notch in sensor cover with the

raised keyway on sensor socket holder (see figure below).  

Note:  Do  not  force  sensor  into  socket.   Misalignment  can

damage the sensor pins.

IMPORTANT:  Make  sure  sensor  is  fully  inserted  for  proper 

operation. The detector will stay in warm-up mode and the on/off 

switch will become inoperative if the sensor becomes loose or is 

not fully inserted.

(do not twist to remove)

Tracer Gas Detection 

Tracer Gas consists of a non-toxic, non-flammable, 

environmentally friendly (non-polluting) mixture of 5% 

hydrogen and 95% nitrogen. For this reason, the Tracer 

Gas can be released into the environment after the leak 

detection procedure is completed. The Tracer Gas mixture 

complies with Article 6, Paragraph 3 of EU directive 

2006/40/EC. 

The H

2

 Leak Detector detects the hydrogen component of 

the tracer and because hydrogen molecules are so small, it 

is an ideal gas for leak detection.  

The Tracer Gas is charged into an empty system at a 

pressure of up to but not exceeding the rated system test 

pressure. As Hydrogen is lighter than air, always probe 

slightly above the suspected leak area.  

Once the source of the leak is located and repaired, the 

Tracer Gas can be released and the system can be 

recharged again with refrigerant.  

Product Specifications 

Model  # 

69341

H

,r

o

t

c

e

t

e

D

k

a

e

L

e

m

a

N

2

 Tracer Gas 

Sensitivity  

< 5 ppm 

Sensor Life 

> 300 hours 

Response Time 

Instantaneous 

Power Supply 

4 AA Alkaline batteries 

Battery Life 

6 hours continuous 

Warm up time 

< 20 seconds 

Probe length 

17 inches 

Weight, lbs 

1

1.5 lbs 

Warranty 

18 months  

Compliancy 

RoHS, CE 

7

To order call 1-800-548-1191 or visit www.partdeal.com - [email protected]

Содержание 69341

Страница 1: ...Model 69341 Detects 5 H2 in 95 Nitrogen Tracer Gas H2 Tracer Gas Leak Detector Operations Manual To order call 1 800 548 1191 or visit www partdeal com info partdeal com ...

Страница 2: ...etector will detect leak rates less than 5 ppm This complies with standards EN35422 and EN14624 The Tracer Gas mixture complies with Article 6 Paragraph 3 of EU Directive 2006 40 EC FEATURES H2 sensitivity 5 ppm Uses 4 AA alkaline batteries True mechanical pump 3 adjustable sensitivity levels Automatic calibration and reset to ambient Long life stable sensor Made in USA 18 month warranty CE Certif...

Страница 3: ...e detector will sound a slow beep Warm up is usually less than 20 seconds 3 READY The detector is ready to begin searching for leaks when the the green Norm sensitivity LED turns on and the audio beep increases in frequency 2 Low Battery Indicator Sensitivity Level H2 Leak Detector Control Panel Low battery indicator Sensitivity level indicator Sensitivity adjust Power On Off 3 To order call 1 800...

Страница 4: ... the 4 AA Alkaline batteries when the red LED on the control panel is lit Follow battery installation instructions under Maintenance section Troubleshooting 4 To order call 1 800 548 1191 or visit www partdeal com info partdeal com ...

Страница 5: ...ector comes with a Leak Test Vial that allows the user to make sure the detector is performing properly To test 1 Remove the plastic seal cap on top of the Leak Test Vial by pulling it off see fig below 2 Turn on the detector and allow the unit to complete the warm up cycle 3 Place the sensor close to the small hole in the top of the Leak Test Vial The beep rate should increase indicating that the...

Страница 6: ...er supplied with the leak detector or they can be re ordered from your supplier or distributor Sensor Failure Mode If the sensor is not working correctly the H2 Leak Detector will not come out of the warm up mode If this happens first be sure the sensor is plugged in all the way If that does not correct the situation replace the sensor Sensor Replacement Remove sensor by pulling out of socket Inst...

Страница 7: ...rged into an empty system at a pressure of up to but not exceeding the rated system test pressure As Hydrogen is lighter than air always probe slightly above the suspected leak area Once the source of the leak is located and repaired the Tracer Gas can be released and the system can be recharged again with refrigerant Product Specifications Model 69341 H r o t c e t e D k a e L e m a N 2 Tracer Ga...

Страница 8: ...ce with the following instructions credit replace or repair at our option the defective leak detector provided it is returned within 18 months of the date of sale H2 Leak Detectors have a date of manufacture serial number located on the label on the bottom of the unit Correction in the manner provided above shall constitute a fulfillment of all liabilities with respect to the quality material and ...

Отзывы: