yellow jacket 40812 Скачать руководство пользователя страница 14

14

Turn the unit off. Rotate the hook (1) out of the 

way of the battery door. Unscrew the battery door 

thumbscrew (2) and remove the battery door. Lift 

the battery pack out somewhat to allow access to 

the battery connection (3). Reconnect the battery 

pack. Tuck the battery harness to the side of the 

battery compartment as shown if Figure 6-2 as 

the battery assembly is lowered into the battery 

compartment.

Coin Cell Battery

If the unit is not maintaining the correct time and 

date, follow the instructions above to remove the 

main battery pack and expose the coin cell battery 

through an access hole in the floor of the battery 

compartment. Using a tool such as a small screw-

driver, deflect the battery to the side until a corner 

of the battery springs up slightly. It may be lifted 

out of the battery socket at this point. Reverse this 

procedure to install the fresh battery.

Spare Parts

Spare parts are listed in Table 6-1. See Chapter 

1 for contact information. See the included TITAN 

manifold instructions for manifold-related spare 

parts.

Table 6-1   Spare Parts

UPC#

Description

40820

Battery Holder, ‘AA’ Size, 2x4

40821

System Analyzer Instrument Boot

40823

System Analyzer Battery Door

40824

System Analyzer Bumper, 

Plugs, & Sensor Boots

40827

USB cable

69101

Vacuum Sensor/Cord Assembly

69217

“K-Type” Clamp Probe (10 ft.)

Software Updates

Details related to software updates are available 

online at www.yellowjacket.com/sa or by contact-

ing Ritchie Engineering. See Chapter 1 for contact 

information.

Replacing the Batteries

The main batteries will need to be replaced on 

a routine basis. Battery performance issues are 

discussed in Chapter 2. The coin cell battery 

maintains the time and date information. It may 

need to be replaced once or twice in the life

of the instrument. To exit the Set-up display, press 

the “Menu” key.

Figure 5-6. Battery Access

Figure 5-7. Battery Servicing

Main Batteries

See Chapter 7 for battery type specifications.

Warning

Because this instrument allows for various 

inputs including electrical and mechanical, 

care must be taken to observe any ways that 

an electrical shock hazard could develop. 

Example: Wet or humid conditions, along with 

a damaged thermocouple or vacuum sensor, 

could allow an electrical path across the instru-

ment and over wet hoses. Keep all intercon-

nected equipment clean, organized, and in 

proper condition. Do not use the instrument 

if you are not qualified to recognize potential 

electrical faults.

Содержание 40812

Страница 1: ...ached to the instrument YELLOW JACKET Refrigeration System Analyzer Users Manual UPC 40812 40813 and 40815 Versions 1 06 and higher 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1...

Страница 2: ...ions 15 Safety 15 Chapter Title Pg Features and Specifications 15 Instrumentation Specifications 16 Warranty Policy 16 Table of Contents List of Tables Table Title Pg 1 1 Safety Information 3 1 2 Symb...

Страница 3: ...ced only by Ritchie Engineering Co or authorized service centers Do not use the instrument if it operates ab normally Protection may be impaired When in doubt have the instrument serviced Refer to war...

Страница 4: ...page by page scrolling mode during data log playback Battery Battery connector orientation Chapter 2 Getting Acquainted Introduction This instrument will clearly and accurately report critical inform...

Страница 5: ...is detected Once the data is stored unit will turn off Maximizing Battery Life Battery life decays fastest when the DIGITAL MANIFOLD display is selected the vacuum sen sor is attached and the backlig...

Страница 6: ...Chart Display This display shows the P T Pressure Tempera ture chart for the refrigerant that is selected in the Set Up display Use the or keys to scroll through the chart Press the Menu key to exit...

Страница 7: ...Transducer internal High Side Pressure Transducer internal Temperature Probes The system analyzer will display readings from any K type thermocouple probe with a miniature style connector see picture...

Страница 8: ...erheat measurement the T1 probe should be mea suring the suction line temperature Temp 2 T2 The temperature indicated by the T2 thermocouple For proper subcool ing measurement the T2 probe should be m...

Страница 9: ...High Side 3 When data recording is activated this indicator will gradually change from red to blue indicating the size of one page of data During playback it will help the user to decide whether to mo...

Страница 10: ...for signs of oil before connecting the 69086 vacuum gauge sensor Keep the sensor vertical when possible Connect the gauge directly to the system away from the pump Isolate the pump from the system an...

Страница 11: ...and reporting at a later time Figure 4 3 Data Log Files Erase File Use the arrow keys to highlight the data file to be erased Press clear to erase Erase All Files Press clear to erase all data files...

Страница 12: ...ry default is MM DD YY Time and Date These settings are modified in a slightly different manner Selecting this setting for modification will open a pop up window Each component of the time and date va...

Страница 13: ...as dish soap If the unit encounters enough exposure to liquid chemicals that the chemical wicks between the in strument boot and the case immediately remove the boot and clean the case as described ab...

Страница 14: ...tem Analyzer Instrument Boot 40823 System Analyzer Battery Door 40824 System Analyzer Bumper Plugs Sensor Boots 40827 USB cable 69101 Vacuum Sensor Cord Assembly 69217 K Type Clamp Probe 10 ft Softwar...

Страница 15: ...ls back to past readings Refrigerant Database 98 NIST refrigerant profiles included upgradeable Data Logging Adjustable time intervals from 0 1 to 10 seconds minimum of 24 hours at 1 0 second interval...

Страница 16: ...Company Inc is guaranteed to be free from material and workmanship defects for a period of two years All other YELLOW JACKET products included in this package are guaranteed to be free from material a...

Отзывы: