13
Operation Method
Apparatus/Liquid Sterilization Course
6. when the pressure and temperature reach the set
scope, the running light (sterilizing light) will be on
and the time will start working for effective heat
preservation and sterilization. “Sterilizing Parameters
Table” may be referenced for sterilizing time.
Note: After sterilizing, [End] display, the drain valve shall keep open, from which
traces of steam will be exhausted to discharge the cold air in the pot, and the
sample will be dried and sterilized completely.
7. When countdown for sterilizing finishes, the running
light (finish lamp) will be on. After the buzzer rings a
few times, the sterilizer will stop working automatically,
and the sterilizing operation will complete.
8. Opening the lid: After sterilizing finishes, the power
shall be switched off. As the liquor and utensils, etc.
may spill or be broken, they shall be cooling off
naturally after heating stops, and only after the steam
exhaust valve releases the steam till the pressure
gauge indicates zero, may the upper handle be turned
counterclockwise for the tip of the positioning mark to
point to center of the lower handle in order to open the
lid.
Note: the lid may only be opened after sterilizing finishes, the remaining steam is released
till the pressure gauge indicates zero.