![Yamato SK 100C Скачать руководство пользователя страница 14](http://html.mh-extra.com/html/yamato/sk-100c/sk-100c_instruction-manual_905303014.webp)
12
Operation Method
Apparatus/Liquid Sterilization Course
Follow the procedures below for the setting of apparatus sterilization and liquid sterilization courses.
1. Press key to shift the function light to the
position of sterilizing temperature, and now the
display window will indicate the temperature mode.
The sterilizing temperature may be set as required.
2. Press
▲▼
key to raise the temperature and press
to lower it. After adjusting to the required
temperature, press another time to confirm the new
setting and complete the temperature setting
procedures.
3. press key to shift the function lamp to the
position of sterilizing time, and the time mode will be
displayed. The sterilizing time may be set as required.
4. Press
▲
key to increase the time and press
▼
key to reduce it. After adjusting to the required time,
press key another time to confirm the new
setting and complete the time setting procedures. The
sterilizing time may be set with reference to “Sterilizing
Parameters”
5. the running light being on indicates the heater is
heating, and with the rise of temperature inside the
evaporation pot, the cold air inside the pot will be
released from the vent automatically. When the
pressure inside the pot
≧
0.028Mpa, the safety lock
will close automatically to lock the upper and lower
handles for safety, and now the hand in the pressure
gauge and the digits in the display window will go up
gradually with heating to indicate the pressure and
temperature inside the pot.
Note
:The outside surface of this unit may get hot to the touch contact between unprotected skin and
outsidesurface of this unit during operation may cause burns.Always use protective gear to shield operator
from hot surfaces.