14
2. Before operating the unit
Installation procedures
8.Connect the membrane filter firmly.
●
Install the membrane filter as follows.
●
Connect firmly. Otherwise, water may leak through the threaded portion.
(1) Select the membrane filter and seal tape from accessories of the main body.
(2) Pay attention to the arrow direction of membrane filter and wind the seal tape, clockwise
as viewed from the winding side, by two to three turns while giving slight tension. Cut
off any surplus tape. In this case, start winding of the seal tape while keeping about two
leading threads free.
(3) With the seal-taped portion on top, screw in the membrane while taking care not to
damage threads. Check for leakage in the course of sampling of pure water. If not,
screw in further. The rough guidance of tightening number is two and a half to three
turns.
※
Be sure to securely store the seal tape, which be necessary when replacing the membrane
filter.
Socket
Form water inlet on
the bask of the main
unit
Membrane filter
Seal tape
Содержание Pure Line WL320A
Страница 2: ......
Страница 44: ...41 11 Electric Circuit Diagram Wiring diagram ...