![Yamato DP-6700 Скачать руководство пользователя страница 39](http://html2.mh-extra.com/html/yamato/dp-6700/dp-6700_operation-manual_3164713039.webp)
37
Display Cause Troubleshooting
The total overflow
The accumulated weight
or count exceeds the
memory capacity.
Reset the total weight and total count
after confirming their data.
Trouble in electronic circuit
This message appears
when a trouble was
detected in the electronic
circuit.
Turn off the scale. And reapply the
power after a brief interval.
When the scale does not return to
normal, contact the dealer.
Trouble in electronic circuit
This message appears
when a trouble was
detected in the electronic
circuit.
Turn off the scale. And reapply the
power after a brief interval.
When the scale does not return to
normal, contact the dealer.
Calibaration error
The weights used for the
calibration may be
incorrect.
Re-calibrate the scale with the proper
weights.
When the scale does not return to
normal, contact the dealer.
Trouble in electronic circuit
This message appears
when a trouble was
detected in the electronic
circuit.
Turn off the scale. And reapply the
power after a brief interval.
When the scale does not return to
normal, contact the dealer.
Trouble in electronic circuit
This message appears
when a trouble was
detected in the electronic
circuit.
Turn off the scale. And reapply the
power after a brief interval.
When the scale does not return to
normal, contact the dealer.