36
Display Cause Troubleshooting
The initial error (B)
Turned on the scale while
a foreign material was
stuck between the main
body and the platform
cover.
Check if a foreign material is not stuck or
does not contact the platform cover.
Then press
. When the scale does
not return to normal, contact the dealer
from which the scale was purchased for
troubleshooting.
The initial error (C)
Pressed
while a
heavy item was on the
platform during the
operation.
The weight of the item is beyond the
level where
is operative.
Remove the item from the platform,
and then press
.
The initial error (D)
Turned on the scale while
the item was left on the
platform. And then,
removed it and pressed
.
Make sure that nothing is left on the
platform and turn on the scale again.
The data overflow error
The stored data exceeds
the memory capacity.
Clear the stored data by sending out.
The printer error
A roll paper is not set or
the printer lid is open.
Make sure that a roll paper is set and
the printer lid is closed. And then turn
the scale again.
The communication error
The communication
program or network is not
active or the distance is too
far.
Make sure that the program and
network are active and locate the
scale closer. And then turn the scale
again.
The USB overflow alarm
The stored data to be sent
to USB exceeds 90% of the
memory capacity.
Clear the stored data by sending to
USB.
Содержание DP-6700