COUNTING FUNCTION
PRE-SETTING
Operation
Display
1 Press to turn on the scale.
2 While pressing , press
to enter the user parameter setting
mode.
Set the parameter # 01 at 004 (Counting function) by pressing
or
.
SETUP
Calculate the average unit weight from 100 samples.
Operation
Display
1 Press to turn on the scale.
Press
again to go to the setting stage of sampling number.
2
Press
or
to set the sampling number at 100.
moves the number 10
→
20
→
50
→
100
→
200
moves the number 200
→
100
→
50
→
20
→
10
Press
after setting the sampling number.
3 Place on the platform 100 pieces of samples
The judgment lamp flickers in blue when the calculated unit weight is
heavier than the minimum division.
The judgment lamp flickers in yellow when the calculated unit weight is
lighter than the minimum division. In this case the risk for generating
weighing (counting) error may be bigger.
4
Press
again to store the unit weight calculated from 100 samples.