23
Error code
Panel display
Conditions that trigger the error code
Corrective action
When there are any objects placed on the
platform of the instrument, pressing the
key to turn ON the display may
generate this code.
Remove the object from the
platform and press the
key.
The panel shows “0.00”;
and the instrument is
operational.
When there are any objects placed on the
platform of the instrument, pressing the
key to perform the zero reset may
generate this code.
When foreign material is caught between
the platform of the instrument and the
container or when the platform is
removed, pressing the
key to turn
ON the display may generate this code.
Correctly install the
platform and press the
key.
The panel shows “0.00”;
and the instrument is
operational.
When the platform of the instrument is
removed, pressing the
key to
perform the zero reset may generate this
code.
Placing any objects that are heavier than
the weighing capacity of the instrument
displays this sign during operation.
Removing the object
returns to the weighing
display. Use the instrument
within its rated operating
range.
Removing the platform of the instrument
displays this sign during operation.
Correctly install the
platform.
The panel shows “0.00”;
and the instrument is
operational.
When the number of measurement times
added has reached the maximum number
(9999), further addition of the number
generates this code.
Press the
key. Then
clear the number of
measurement times added
by referring to page 15.
This code appears when the added value
exceeds the maximum value
(999,999,999).
Press the
key to reset
the error. Then clear the
added value by referring to
page 15.
This code appears in the event of
abnormal A/D conversion., or when the
connected optional journal printer does
not function properly due to a paper jam,
etc.
Check the printing section
for the paper jam, etc.
Power OFF and then ON
by pressing the
key
and then the
key.
<When the optional RS232C is in use>
This code appears when connected
equipment is not ready for receiving a
signal.
Check the connector
connection. Press the
key to reset the error
and
then restart
communication.
Содержание DP-5601
Страница 8: ...6 Display and operation key sections ...
Страница 31: ...29 Desk top Type Separate Stand Type ...