![background image](http://html.mh-extra.com/html/yamato/dnf811/dnf811_instruction-manual_905327028.webp)
24
4. Operating procedure
Fixed temperature operation
6 Stop running Fixed Temperature
Operations with timer setting.
(
Quick Automatic Stop Function
)
Quick Automatic Stop Function is to stop
automatically running Fixed Temperature
Operation.
①
Press
key at running Fixed Temperature
operation.
②
Show [
QSTOP]
on Bottom Screen and start [TIME]
lamp flashing on the left top of Bottom Screen.
③
Select the method to stop from TIME/CLOCK with
key and then press
key.
④
Set TIME
(
capable setting range: 0
~
99hr
:
59min
)
or CLOCK (according to 24-hour time) on
Top Screen and then press
key.
Example 1.
Setting time to stop:
Operation is stopped automatically in 35 hours and 30
minutes
once
temperature reached
to
target
temperature.
Example 2.
Setting clock time to stop:
Operation is stopped automatically at 15:00.
⑤
The AUTO STOP (Auto Stop) lamp comes on and
the Auto Stop function starts.
※
You can use the
key to check the remaining
operation time/stop time information on the Bottom
screen.
※
Screen to check the remaining operation time
※
Screen to check the operation stop time
Press the
key again or wait for about 10
seconds to return to the original status.