Indexes
VARiON
®
Plus
700 IQ H
56
ba76207e02
12/2020
10 Indexes
10.1 Explanation of the messages
This chapter contains a list of all the message codes and related
message texts that can occur in the log book of the IQ
S
ENSOR
N
ET
system for the VARiON
®
Plus
700 IQ H sensor.
10.1.1 Error messages
Information on
the contents and structure of the log book and
the structure of the message code
is given in the
L
OG
BOOK
chapter of the IQ
S
ENSOR
N
ET
sys-
tem operating manual.
The last three digits of the message code identify the
source of the message:
521 =
700IQ
(Armature / component class,
adapters ADA)
395 =
VARiON A
(Ammonium / potassium electrode)
396 =
VARiON N
(Nitrate / chloride sensor)
397 =
VARiON A
(Potassium electrode)
398 =
VARiON N
(Nitrate sensor)
Message code
Message text
EA1395
Meas. range exceeded or undercut
* Check process
* Select other meas. range
EA1396
Meas. range exceeded or undercut
* Check process
* Select other meas. range
EA2521
Sensor temperature too high!
* Check process and application
EA3521
Sensor temperature too low!
* Check process and application
EAN395
Potassium measurement: range exceeded or undercut
* Check process
Содержание YSI VARiON Plus 700 IQ H