Commissioning
VARiON
®
Plus
700 IQ H
24
ba76207e02
12/2020
Measuring range
(
Measuring mode
:
mV
)
-2000 ... 2000 mV
Fixed range
If the sensor is not equipped with a chloride electrode:
Chloride compens.
Manual
After determining the chloride content of
the test solution enter the chloride content
determined in the next line manually
(
Chloride conc.
).
The measured value is accordingly
corrected by the entered chloride content.
Note: With the
Link (sensor to sensor)
function, the IQ
S
ENSOR
N
ET
is able to
provide a measured value for the
interfering ion.
Note: Detailed information on the subject
of chloride compensation is given in the
primer,
I
ON
SELECTIVE
MEASUREMENT
IN
ONLINE
ANALYSIS
.
Chloride conc.
0.1 ... 1000 mg/l
Manual input of the chloride value
If the sensor is equipped with a chloride electrode:
VARiON Cl
hide
show
hide
(standard setting):
The chloride electrode is not displayed as
an extra sensor in the measured value
display.
show
:
If the sensor is equipped with the chloride
electrode, the electrode is displayed as the
extra sensor
VARiON Cl
in the measured
value display.
The relevant settings can be done in an
extra setting table (see section
Note: If the maximum number of sensors
for the system is exceeded by activating
the sensor, the
VARiON Cl
cannot be acti-
vated.
Menu item
Selection/Values
Explanations