VARiON
®
Plus
700 IQ H
Maintenance and electrode exchange
43
ba76207e02
12/2020
6.4
Polishing the chloride electrode
Caused by the test medium, a coating can develop on the surface of
the chloride electrode, which gradually reduces the electrode slope. To
maintain a proper chloride compensation the surface must be renewed
by polishing at regular intervals.
The polishing can be done while the electrode is mounted (remove the
protective hood as necessary). For polishing, exclusively use the SF
300 polishing strip included in the scope of delivery of the chloride elec-
trode (see chapter 7 R
). Moisten
the rough side of the polishing strip with water and polish off any coat-
ings under slight pressure.
Fig. 6-2
Polishing the chloride electrode.
For the correct storage of the electrodes please follow the
instructions in section 3.2
.
NOTICE
Danger of damaging the electrode. Exclusively use the
SF 300 polishing strip. Never use commercial sandpaper
or similar..
SF 300