Setting table
VARiON A
(ammonium sensor)
Menu item
Selection/Values
Explanations
Measuring mode
NH4-N
NH4
mV
Citation form of the mass concentration or
voltage of the electrode.
Measuring range
(
Measuring mode
:
NH4-
N
)
AutoRange
0.1 ... 100.0 mg/l
1 ... 2000 mg/l
2 measuring ranges can be selected.
With
AutoRange
, the instrument
automatically switches to the suitable
measuring range.
Measuring range
(
Measuring mode
:
NH4
)
AutoRange
0.1 ... 129.0 mg/l
1 ... 2580 mg/l
2 measuring ranges can be selected.
With
AutoRange
, the instrument
automatically switches to the suitable
measuring range.
Measuring range
(
Measuring mode
:
mV
)
-2000 ... 2000 mV
Fixed range
If the sensor is not equipped with a potassium electrode:
Potassium compens.
Manual
After determining the potassium content of
the test solution, enter the potassium
content determined in the next line
manually (
Potassium conc.
).
The measured value is accordingly
corrected by the entered potassium
content.
Note: With the
Link (sensor to sensor)
function, the IQ
S
ENSOR
N
ET
is able to
provide a measured value for the
interfering ion.
Note: Detailed information on the subject
of potassium compensation is given in the
primer,
I
ON
SELECTIVE
MEASUREMENT
IN
ONLINE
ANALYSIS
.
Potassium conc.
0.1 ... 1000 mg/l
Manual input of the potassium value