What to do if...
Turb 430 IR/T
48
ba75507d09
08/2021
6
What to do if...
6.1
General errors
Display,
LoBat
Meter does not
react to keystroke
Error message,
Error
0, 8, 16, 16384
6.2
Turbidity
Error message
Measured values
obviously
incorrect
Measured value
display
< 0.01 FNU
Cause
Remedy
– Batteries or battery pack
almost empty.
– Insert new batteries
– Charge the battery pack(see
Cause
Remedy
– Software error
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the
<START/ENTER>
and
<PRT>
key simultaneously.
Cause
Remedy
– Instrument error
– Repeat measurement
– Meter defective,
Send in the meter for repair, stating
the error number
Cause
Remedy
– Cell not correctly inserted
– Lock the cell into place
– Cell contaminated
– Clean the cell
– Calibration too old
– Carry out calibration
Cause
Remedy
– Calibration defective
– Carry out calibration
– Measured value outside the
measuring range
– Not possible