Operation
Turb 430 IR/T
28
ba75507d09
08/2021
Display with mea-
suring range over-
flow
If the measured value is outside the measuring range of the Turb 430 IR, it is
indicated on the display:
4.6.1
Settings for turbidity measurement
Overview
For turbidity measurements, the following settings are possible in the menu
Konfiguration
/
Trübung
:
Kalibrierung
Rücksetzen
4
Fill the cell with the sample to be measured (approx. 15 ml). Close
the cell with the black light protection cap.
5
Insert the cell (see section 4.2).
6
Align the marking on the cell cap with the marking at the cell shaft.
or
Determine the suitable position of the cell in the cell shaft (see D
7
Press the
<START/ENTER>
key.
Measurement starts. The measured value is displayed.
The measured value is automatically output to the interfac-
es.
8
Repeat the steps 2 to 8 for further samples.
Trübung
157.0
FNU
NTU
[3-P StdCAL]
15.01.20 15:12
Trübung
> 1100
FNU
NTU
[3-P StdCAL]
15.01.20 15:12