If you want to clean and rinse the ground junction, you can open it as follows: 

SenTix H and SenTix HW: completely unscrew the ground stopper (see 
figure on Seite 5).

SenTix HWD: slide the ground sleeve from the ground stopper towards the 
shaft while turning it slightly. 

After cleaning

Rinse the combination electrode with deionized water and condition it in 
reference electrolyte solution for at least 1 hour. Then recalibrate the combi-
nation electrode.

Wear parts and accessories

Note

Detailed information on our wide range of buffer solutions and more acces-
sories is given in the price list of the WTW catalog "Laboratory and field in-
strumentation".

Description

Model

Order no.

Reference electrolyte solution 250 mL
(KCl 3 mol/L, Ag

+

 free)

KCl-250

109 705

Pepsin cleaning solution 3 x 250 ml

PEP/pH

109 648

Connection cable S7 plug-in connector/DIN, 1 m

AS/DIN

108 110

Connection cable S7 plug-in connector/DIN, 3 m

AS/DIN-3

108 112

Connection cable S7 plug-in connector/BNC, 
1 m

AS/BNC

108 114

Содержание WTW SenTix H

Страница 1: ...OPERATING MANUAL ba75739e02 07 2013 SenTix PH COMBIN ELECTRODES WITH LIQUID REFERENCE SYSTEM AND GROUND JUNCTION SenTix H SenTix HW SenTix HWD...

Страница 2: ...opyright Weilheim 2013 WTW GmbH Reproduction in whole or even in part is prohibited without the express written permission of WTW GmbH Weilheim Printed in Germany For the most recent version of the ma...

Страница 3: ...round No Adjustable ground junc tion SenTix HWD KCl 3 mol L Ag free Ground Yes WTW model pH measur ing range Allowed tempera ture range Membrane resis tance at 25 C Typical application SenTix H 0 14 0...

Страница 4: ...for the combination electrode to be stored Always keep the watering cap clean Connect the combination electrode to the meter Calibrate the combination electrode according to the operating manual of th...

Страница 5: ...olyte solu tion comes out Calibration and measurement General rules Make sure the refilling opening for the reference electrolyte solution is open Avoid the displacement of any solution sample or buff...

Страница 6: ...ning Note pH combination electrodes must not be stored dry or in deionized water The combination electrode could be permanently damaged by this If the liquid in the watering cap has dried up condition...

Страница 7: ...the tip of the dropping bottle into the refilling opening while turning it slightly Pump the reference electrolyte in the shaft using the dropping bottle Pull the dropping bottle out of the refilling...

Страница 8: ...lectrolyte solution for at least 1 hour Then recalibrate the combi nation electrode Wear parts and accessories Note Detailed information on our wide range of buffer solutions and more acces sories is...

Страница 9: ......

Страница 10: ......

Страница 11: ......

Страница 12: ...water needs Developing new technologies that will improve the way water is used conserved and re used in the future is central to our work We move treat analyze and return water to the environment and...

Отзывы: