Wear parts and accessories
SenTix
®
8
ba75739e02
07/2013
If you want to clean and rinse the ground junction, you can open it as follows:
SenTix H and SenTix HW: completely unscrew the ground stopper (see
figure on Seite 5).
SenTix HWD: slide the ground sleeve from the ground stopper towards the
shaft while turning it slightly.
After cleaning
Rinse the combination electrode with deionized water and condition it in
reference electrolyte solution for at least 1 hour. Then recalibrate the combi-
nation electrode.
Wear parts and accessories
Note
Detailed information on our wide range of buffer solutions and more acces-
sories is given in the price list of the WTW catalog "Laboratory and field in-
strumentation".
Description
Model
Order no.
Reference electrolyte solution 250 mL
(KCl 3 mol/L, Ag
+
free)
KCl-250
109 705
Pepsin cleaning solution 3 x 250 ml
PEP/pH
109 648
Connection cable S7 plug-in connector/DIN, 1 m
AS/DIN
108 110
Connection cable S7 plug-in connector/DIN, 3 m
AS/DIN-3
108 112
Connection cable S7 plug-in connector/BNC,
1 m
AS/BNC
108 114