Ground junction of

SenTix H and

SenTix HW

With the ground junction of the SenTix H and SenTix HW combination elec-
trodes, the electrode outflow can be adjusted with a thread:

Hold the combination electrode with the pH glass membrane pointing 
downward. Using the provided special wrench, open the ground junction 
until some electrolyte solution can be seen to come out of the junction and 
all ground surfaces are completely moist. 

Carefully and using little power close the junction until the contact between 
the rough ground surfaces can be noticed as a resistance. The combina-
tion electrode is now ready to measure.

Caution: Always use very little power to close the junction so the 
ground surfaces are not damaged! 

The ground junction can be opened further so that more electrolyte solu-
tion comes out. 

Calibration

and measurement:

General rules

Make sure the refilling opening for the reference electrolyte solution is 
open.

Avoid the displacement of any solution (sample or buffer solution) from 
one measurement to the next by taking the following measures:

– Shortly rinse the calibration and sample beakers with the solution the 

beakers are to be filled with next.

– Between measurements, rinse the combination electrode with the so-

lution that follows. Alternatively, you can also rinse the combination 
electrode with deionized water and then carefully dab it dry.

Immerse the combination electrode in the solution in a vertical or slightly 
tilted position.

Make sure the immersion depth is correct. The ground junction must be 
completely submersed in the solution. At the same time, the level of the 
reference electrolyte must be at least 2 cm above the level of the solution.

Provide approximately the same stirring conditions for measuring as for 
calibrating.

S

pe

c

i

al wr

en

ch

Gr

o

u

n

d ju

n

ct

ion

O

pen

Cl

o

s

e

Содержание WTW SenTix H

Страница 1: ...OPERATING MANUAL ba75739e02 07 2013 SenTix PH COMBIN ELECTRODES WITH LIQUID REFERENCE SYSTEM AND GROUND JUNCTION SenTix H SenTix HW SenTix HWD...

Страница 2: ...opyright Weilheim 2013 WTW GmbH Reproduction in whole or even in part is prohibited without the express written permission of WTW GmbH Weilheim Printed in Germany For the most recent version of the ma...

Страница 3: ...round No Adjustable ground junc tion SenTix HWD KCl 3 mol L Ag free Ground Yes WTW model pH measur ing range Allowed tempera ture range Membrane resis tance at 25 C Typical application SenTix H 0 14 0...

Страница 4: ...for the combination electrode to be stored Always keep the watering cap clean Connect the combination electrode to the meter Calibrate the combination electrode according to the operating manual of th...

Страница 5: ...olyte solu tion comes out Calibration and measurement General rules Make sure the refilling opening for the reference electrolyte solution is open Avoid the displacement of any solution sample or buff...

Страница 6: ...ning Note pH combination electrodes must not be stored dry or in deionized water The combination electrode could be permanently damaged by this If the liquid in the watering cap has dried up condition...

Страница 7: ...the tip of the dropping bottle into the refilling opening while turning it slightly Pump the reference electrolyte in the shaft using the dropping bottle Pull the dropping bottle out of the refilling...

Страница 8: ...lectrolyte solution for at least 1 hour Then recalibrate the combi nation electrode Wear parts and accessories Note Detailed information on our wide range of buffer solutions and more acces sories is...

Страница 9: ......

Страница 10: ......

Страница 11: ......

Страница 12: ...water needs Developing new technologies that will improve the way water is used conserved and re used in the future is central to our work We move treat analyze and return water to the environment and...

Отзывы: