SenTix

®

Maintenance and cleaning

7

ba75739e02

07/2013

Maintenance and cleaning

Refilling the

reference

electrolyte

During operation, a small amount of reference electrolyte leaks through the 
junction from the combination electrode into the test sample. If the level of ref-
erence electrolyte becomes too low with time, refill it through the refilling 
opening. Refilling is very easy with the enclosed dropping bottle. Proceed as 
follows:

Cut off the tip of the dropping bottle at a right angle until the opening in the 
tip can be seen

Open the refilling opening of the combination electrode

Press the tip of the dropping bottle into the refilling opening while turning 
it slightly

Pump the reference electrolyte in the shaft using the dropping bottle

Pull the dropping bottle out of the refilling opening while turning it slightly 
if necessary.

Cleaning

Remove water-soluble contamination by rinsing with deionized water. Re-
move other contamination as follows:

Note

Hydrofluoric acid, hot phosphoric acid and strong alkaline solutions destroy 
the glass membrane.

Contamination

Cleaning procedure

Fat and oil

Rinse with water containing household 
washing-up liquid

Lime and hydroxide deposits Rinse with citric acid (10 % by weight)

Proteins

Immerse in pepsin cleaning solution PEP/pH 
for approx. 1 hour. 
Note: Make sure the level of the reference 
electrolyte is above that of the cleaning solu-
tion.

Содержание WTW SenTix H

Страница 1: ...OPERATING MANUAL ba75739e02 07 2013 SenTix PH COMBIN ELECTRODES WITH LIQUID REFERENCE SYSTEM AND GROUND JUNCTION SenTix H SenTix HW SenTix HWD...

Страница 2: ...opyright Weilheim 2013 WTW GmbH Reproduction in whole or even in part is prohibited without the express written permission of WTW GmbH Weilheim Printed in Germany For the most recent version of the ma...

Страница 3: ...round No Adjustable ground junc tion SenTix HWD KCl 3 mol L Ag free Ground Yes WTW model pH measur ing range Allowed tempera ture range Membrane resis tance at 25 C Typical application SenTix H 0 14 0...

Страница 4: ...for the combination electrode to be stored Always keep the watering cap clean Connect the combination electrode to the meter Calibrate the combination electrode according to the operating manual of th...

Страница 5: ...olyte solu tion comes out Calibration and measurement General rules Make sure the refilling opening for the reference electrolyte solution is open Avoid the displacement of any solution sample or buff...

Страница 6: ...ning Note pH combination electrodes must not be stored dry or in deionized water The combination electrode could be permanently damaged by this If the liquid in the watering cap has dried up condition...

Страница 7: ...the tip of the dropping bottle into the refilling opening while turning it slightly Pump the reference electrolyte in the shaft using the dropping bottle Pull the dropping bottle out of the refilling...

Страница 8: ...lectrolyte solution for at least 1 hour Then recalibrate the combi nation electrode Wear parts and accessories Note Detailed information on our wide range of buffer solutions and more acces sories is...

Страница 9: ......

Страница 10: ......

Страница 11: ......

Страница 12: ...water needs Developing new technologies that will improve the way water is used conserved and re used in the future is central to our work We move treat analyze and return water to the environment and...

Отзывы: