Maintenance, cleaning, disposal, and replacement
TriOxmatic
®
700 (SW)
70
ba25307e06
07/2015
5.4.2
Cleaning the silver counter electrode
We recommend to clean the silver counter electrode with the RA 600
cleaning attachment that is available as an accessory (see section
5.8 M
). This prevents
the reference electrode from coming into contact with the cleaning
solution. However, it is also possible to clean the silver counter
electrode in a beaker.
Caution
The reference electrode must not come into contact with the
cleaning solution under any circumstances. This could destroy
the reference electrode and cause the sensor to become
defective. However, no danger can arise if the RA 600 is used
correctly.
Variant 1:
Cleaning the silver
counter electrode with
the RA 600
cleaning attachment
Caution
Never lubricate with oil or grease the inner O-ring in the RA 600
cleaning attachment!
1
Wipe the silver counter electrode with a lint free paper towel
and carefully remove any loose deposits.
2
Screw the cleaning attachment onto the sensor, instead of the
membrane cap.