TriOxmatic
®
700 (SW)
Measuring / Operation
53
ba25307e06
07/2015
4
Measuring / Operation
4.1
Measuring
Warning
Contact with the sample can lead to danger to the user!
Depending on the type of sample, suitable protective measures
must be taken (protective clothing, protective goggles, etc.).
For measuring submerse the operable sensor in the test sample. The
measured value is available immediately on submersing.
Note
Do not suspend the sensor on the sensor connection cable. Use a
sensor holder or armature. Information on this and other
TriOxmatic
®
700 (SW) accessories is given in the WTW catalog and on
the Internet.
Minimum approach flow
The minimum required approach flow at the sensor must be present
during measurement (see chapter 7 T
approach flow can, e.g. be present due to:
natural movement of the water (flow rate, sea disturbance)
turbulences in the activated sludge basin.