Settings
Multi 3630 IDS
88
ba77170e03
10/2017
11.7.1 Resetting the measurement settings
pH
The following settings for pH measurements are reset to the default settings
with the
Reset
function:
The sensor settings are reset under the
Reset
menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the
<ENTER>
key.
ORP
The following settings for ORP measurements are reset to the default settings
with the
Reset
function:
The sensor settings are reset under the
Reset
menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the
<ENTER>
key.
Dissolved oxygen
The following settings for D.O. measurements are reset to the default settings
with the
Reset
function:
The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
Setting
Default settings
Buffer AutoCalTEC
Calibration interval
7 d
Unit for slope
mV/pH
Measured parameter
pH
pH resolution
0.001
mV resolution
0.1
Asymmetry
0 mV
Slope
-59.2 mV
Man. temperature
25 °C
Single-point calibration
off
Setting
Default settings
mV resolution
0.1
Man. temperature
25 °C
Содержание WTW OxiTop-IDS Multi 3630 IDS
Страница 1: ...OPERATING MANUAL ba77170e03 10 2017 Multi 3630 IDS DIGITAL METER FOR IDS SENSORS 3630 IDS...
Страница 2: ...Multi 3630 IDS 2 ba77170e03 10 2017 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Страница 15: ...Multi 3630 IDS Commissioning ba77170e03 10 2017 15 3 3 4 Setting the date and time 3 See section 4 5 5...
Страница 117: ......