Multi 3630 IDS
Settings
ba77170e03
10/2017
83
Setting
menuTetraCon 925
Menu item
Possible
setting
Explanation
Calibration
/
Calibration record
-
Displays the calibration record of the
last calibration.
Calibration
/
Calibration data
storage
/
Display
-
Displays the last calibration records
(max. 10)
Calibration
/
Calibration data
storage
/
Output to
USB flash drive or
printer
-
Outputs the stored calibration data
to a connected USB memory device/
USB printer)
Calibration
/
Calibration data
storage
/
Output to
RS232/USB
-
Outputs the calibration data storage
to the interface
Calibration
/
Calibration interval
1 ...
150
...
999 d
Calibration interval
for the IDS con-
ductivity sensor (in days).
The meter reminds you to calibrate
regularly by the flashing sensor sym-
bol in the measuring screen.
Type
Measuring cell used
Cal
Measuring cells whose cell constant
is determined by calibration in the
KCL control standard solution.
Calibration range:
0.450 to 0.500 cm
-1
The currently valid cell constant is
displayed in the status line.
man
Freely adjustable cell constant in the
range 0.450 ... 0.500 cm
-1
.
Man. cell const.
0.450 ...
0.475
...
0.500 cm
-1
Display and setting options for the
cell constant.
Temp. comp. (TC)
/
Method
nLF
Lin
off
Procedure for temperature compen-
sation (see section 8.2).
This setting is only available for the
measured parameters,
ϰ
and
ρ
.
Temp. comp. (TC)
/
Linear coeff.
0.000 ...
2.000
...
10.000 %/K
Coefficient of the linear temperature
compensation.
This menu item is only available
when the linear temperature com-
pensation is set.