![Xylem WTW OxiTop-IDS Multi 3630 IDS Скачать руководство пользователя страница 84](http://html1.mh-extra.com/html/xylem/wtw-oxitop-ids-multi-3630-ids/wtw-oxitop-ids-multi-3630-ids_operating-manual_4033885084.webp)
Settings
Multi 3630 IDS
84
ba77170e03
10/2017
Setting menu
LR 925/01
11.5 Turb measurement settings
11.5.1 Settings for turbidity sensors
The settings are made in the menu for the measured parameter, turbidity. To
open the settings, display the required measured parameter in the measured
value display and press the
<ENTER>
key. After completing the settings,
switch to the measured value display with
<M>
.
The possible settings are individually displayed for each sensor. Default set-
tings are printed in
bold
.
Temp. comp. (TC)
/
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available for the
measured parameters,
ϰ
and
ρ
.
TDS factor
0.40 ...
1.00
Factor for TDS value
Reset
-
Resets all sensor settings to the
delivery condition (see section
11.7.1).
Menu item
Possible
setting
Explanation
Menu item
Possible
setting
Explanation
Cell constant
0.090 ... ...
0.100
...
0.110 cm
-1
Display and setting options for the
cell constant
Temp. comp. (TC)
/
Method
nLF
Lin
off
Procedure for temperature com-
pensation (see section 8.2).
This setting is only available for the
measured parameters,
ϰ
and
ρ
.
Temp. comp. (TC)
/
Linear coeff.
0.000 ...
2.000
...
10.000 %/K
Coefficient of the linear tempera-
ture compensation.
This menu item is only available
when the linear temperature com-
pensation is set.
Temp. comp. (TC)
/
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available for the
measured parameters,
ϰ
and
ρ
.
TDS factor
0.40 ...
1.00
Factor for TDS value
Reset
-
Resets all sensor settings to the
delivery condition (see section
11.7.1).