Operation
Oxi 3210
36
ba75795e07
10/2021
Editing the data memory
The memory is edited in the menu,
Storage & config
/
Data storage
. To
open the
Storage & config
menu, press the
<F1__>
/[Menu] key in the
measured value display.Open the manual memory directly with the
<RCL>
key.
Settings
Display presentation of
a dataset
Quitting the display
To quit the display of stored measurement datasets, you have the fol-
lowing options:
Switch directly to the measured value display with
<M>
.
Quit the display and move to the next higher menu level with
<F1>
/
[Back]
.
Menu item
Setting/
function
Explanation
Data storage
/
Manual data storage
/
Display
-
Displays all measurement
datasets page by page.
Further options:
Scroll through the data-
sets with
<
><
>
.
Quit the display with
<F1>
/
[Back]
.
Data storage
/
Manual data storage
/
Erase
-
Erases the entire manual
measurement data memory.
Note:
All calibration data remain
stored when this action is
performed.
Manual data storage
3 of 64
15.03.2014 11:24:16 ID number: 1
O2 7.92 mg/l 25.1 °C AR +++
Sal 6.5
15.03.2014
08:00
Back
Содержание WTW Oxi 3210
Страница 1: ...OPERATING MANUAL ba75795e07 10 2021 Oxi 3210 DISSOLVED OXYGEN D O METER...
Страница 2: ...2 ba75795e07 10 2021 Oxi 3210 Copyright 2021 Xylem Analytics Germany GmbH Printed in Germany...
Страница 5: ...Oxi 3210 Contents 5 ba75795d07 10 2021...
Страница 10: ...Overview Oxi 3210 10 ba75795e07 10 2021...
Страница 14: ...Safety Oxi 3210 14 ba75795e07 10 2021...
Страница 41: ...Oxi 3210 Maintenance cleaning disposal 41 ba75795e07 10 2021...
Страница 43: ...Oxi 3210 What to do if 43 ba75795e07 10 2021...
Страница 46: ...Technical data Oxi 3210 46 ba75795e07 10 2021...
Страница 50: ...Lists Oxi 3210 50 ba75795d07 10 2021...
Страница 51: ......