Criteria for a stable
measured value
The
Stability control
function checks whether the measured values are
stable within the monitored time interval.
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
4.4.3
Settings for D.O. sensors (menu for measurement and cal-
ibration settings)
Overview
The following settings are possible for D.O. sensors:
Salinity correction
Salinity (salinity equivalent)
Calibration interval
Automatic
Stability control
Settings
The settings are available in the menu for measurement and calibration
settings. To open the settings, display the required parameter in the
measured value display and press the
<F1>
/
[Menu]
or
<ENTER>
key.
After completing the settings, switch to the measured value display with
<M>
.
Measured parameter
Time interval
Stability in the time inter-
val
D.O. concentration
20 seconds
Δ
: better than 0.05 mg/l
D.O. saturation
20 seconds
Δ
: better than 0.6 %
D.O. partial pressure
20 seconds
Δ
: Better than 1.2 mbar
Temperature
15 seconds
Δ
: better than 0.5 °C
Menu item
Possible
setting
Explanation
Calibration
/
Calibration record
-
Displays the calibration
record of the last calibration.
Calibration
/
Calibra-
tion data storage
-
Displays the last calibration
records.
Содержание WTW Oxi 3210
Страница 1: ...OPERATING MANUAL ba75795e07 10 2021 Oxi 3210 DISSOLVED OXYGEN D O METER...
Страница 2: ...2 ba75795e07 10 2021 Oxi 3210 Copyright 2021 Xylem Analytics Germany GmbH Printed in Germany...
Страница 5: ...Oxi 3210 Contents 5 ba75795d07 10 2021...
Страница 10: ...Overview Oxi 3210 10 ba75795e07 10 2021...
Страница 14: ...Safety Oxi 3210 14 ba75795e07 10 2021...
Страница 41: ...Oxi 3210 Maintenance cleaning disposal 41 ba75795e07 10 2021...
Страница 43: ...Oxi 3210 What to do if 43 ba75795e07 10 2021...
Страница 46: ...Technical data Oxi 3210 46 ba75795e07 10 2021...
Страница 50: ...Lists Oxi 3210 50 ba75795d07 10 2021...
Страница 51: ......