What to do if...
Multi 3320
90
ba77053e04
07/2018
14.4 Conductivity
Error message
OFL, UFL
The measured value is outside the measuring range.
Error message,
Error
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause
Remedy
– The measured value is outside
the measuring range of the meter
– Use a suitable measuring cell
Cause
Remedy
– Measuring cell contaminated
– Clean cell and replace it if nec-
essary
– Calibration solution not suitable
– Check the calibration solutions