Conductivity
Multi 3320
52
ba77053e04
07/2018
9
Conductivity
9.1
Measuring
9.1.1
Measuring the conductivity
NOTE
When connecting an grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect
.
The USB interface is not galvanically isolated.
1.
Connect the conductivity measuring cell to the measuring instrument.
The conductivity measuring screen is displayed.
2.
If necessary, select the measured parameter with
<M>
.
3.
Check the
Measuring cell
settings and cell constant for the connected
conductivity measuring cell.
If necessary, correct the settings.
The selection of the measuring cell and the setting of the cell con-
stant is done in the measurement settings menu for conductivity
(see section 10.1.8 S
page 66). The cell constant to be set must either be taken from the
operating manual of the measuring cell or is printed on the measur-
ing cell.
4.
Immerse the conductivity measuring cell in the test sample.
5.
If necessary, select the measured parameter with
<M>
.
Conductivity [µS/cm] / [mS/cm]
Resistivity [
Ω
·cm] / [k
Ω
·cm] / [M
Ω
·cm]
Salinity SaL [ ]
Total dissolved solids TDS [mg/l] / [g/l]
ϰ
1417
µS/cm
25.0
°C
[Tr25][nLF][
0.470 1/cm
]
USB output
15.02.2014
08:00
Содержание WTW Multi 3320
Страница 1: ...OPERATING MANUAL ba77053d04 07 2018 Multi 3320 METER FOR 2 SENSORS PH ORP D O COND 3320 ...
Страница 8: ...Overview Multi 3320 8 ba77053e04 07 2018 ...
Страница 102: ...Glossary Multi 3320 102 ba77053e04 07 2018 ...
Страница 104: ...Index Multi 3320 104 ba77053e04 07 2018 Two point calibration ISE 42 pH 26 30 Z Zero point of pH electrode 25 ...
Страница 105: ......