Multi 3320
Dissolved oxygen
ba77053e04
07/2018
51
Calibration record
(example)
Display
Calibration
record
Relative slope
+++
S = 0.8 ... 1.25
++
S = 0.7 ... 0.8
+
S = 0.6 ... 0.7
Error
Error
S < 0.6 or S > 1.25
Error elimination (see section 14
W
Multi 3320
Ser. no. 11292113
CALIBRATION CellOx 325
15.02.2014 16:13:33
S = 0.88
25.0 °C
Sensor
+++
Set 1 for the type, "CellOx":CellOx 325
Set 2, for the type "DurOx":
DurOx 325