CellOx 325
Technical data
17
ba25301e06
12/2017
Pin assignment
Measurement
conditions
Measuring ranges at 20 °C
(68 °F)
0 ... 50 mg/l D.O.
0 ... 600 % D.O. saturation
0 ... 1250 mbar D.O. partial pressure
Temperature range
0 ... 50 °C (32 ... 122 °F)
Max. allowed
overpressure
6·10
5
Pa (6 bar)
Depth of immersion
min. 6 cm
max. 20 m (depending on the cable length)
Operating position
Any
Approach flow
> 3 cm/s at 10 % measurement accuracy
10 cm/s at 5 % measurement accuracy
18 cm/s at 1 % measurement accuracy
Storage
conditions
Recommended storing
method
In the calibration and storage vessel,
OxiCal
®
-SL (moist)
Storage temperature
0 ... 50 °C (32 ... 122 °F)
Characteristics
when delivered
Zero signal
< 0.1 % of the saturation value
Response time at 20 °C
(68 °F)
t
90
(90% of the final value display after) < 10 s
t
95
(95 % of the final value display after) < 16 s
t
99
(99 % of the final value display after) < 60 s
Own consumption of
oxygen at 20 °C (68 °F)
0.008
μ
g·h
-1
(mg/l)
-1
Drift
Approx. 3 % per month in the operating
condition
Response time of
temperature measurement
t
99
(99 % of the final value display after) < 60 s
Precision of temperature
measurement
± 0.2 K
Working life
min. 6 months with one electrolyte filling